116
IRUS Total
Downloads
  Altmetric

Chip-scale coils for millimeter-sized bio-implants

File Description SizeFormat 
08416741.pdfPublished version2.4 MBAdobe PDFView/Open
Title: Chip-scale coils for millimeter-sized bio-implants
Authors: Feng, P
Yeon, P
Cheng, Y
Ghovanloo, M
Constandinou, TG
Item Type: Journal Article
Abstract: Next generation implantable neural interfaces are targeting devices with mm-scale form factors that are freely floating and completely wireless. Scalability to more recording (or stimulation) channels will be achieved through distributing multiple devices, instead of the current approach that uses a single centralized implant wired to individual electrodes or arrays. In this way, challenges associated with tethers, micromotion and reliability of wiring is mitigated. This concept is now being applied to both central and peripheral nervous system interfaces. One key requirement, however, is to maximize SAR-constrained achievable wireless power transfer efficiency (PTE) of these inductive links with mm-sized receivers. Chip-scale coil structures for microsystem integration that can provide efficient near-field coupling are investigated. We develop near-optimal geometries for three specific coil structures: “in-CMOS”, “above-CMOS” (planar coil post-fabricated on a substrate) and “around-CMOS” (helical wirewound coil around substrate). We develop analytical and simulation models that have been validated in air and biological tissues by fabrications and experimentally measurements. Specifically, we prototype structures that are constrained to a 4mm x 4mm silicon substrate i.e. the planar in-/above-CMOS coils have outer diameter <4mm, whereas the around-CMOS coil has inner diameter of 4mm. The in-CMOS and above-CMOS coils have metal film thicknesses of 3μm aluminium and 25μm gold, respectively, whereas the around-CMOS coil is fabricated by winding a 25μm gold bonding-wire around the substrate. The measured quality factors (Q) of the mm-scale Rx coils are 10.5 @450.3MHz (in-CMOS), 24.61 @85MHz (above-CMOS), and 26.23 @283MHz (around-CMOS). Also, PTE of 2-coil links based on three types of chip-scale coils is measured in air and tissue environment to demonstrate tissue loss for bio-implants. The SAR-constrained maximum PTE are measured (together with resonant frequencies, in tissue) 1.64% @355.8MHz (in-CMOS), 2.09% @82.9MHz (above-CMOS), and 3.05% @318.8MHz (around-CMOS).
Issue Date: 1-Oct-2018
Date of Acceptance: 28-Jun-2018
URI: http://hdl.handle.net/10044/1/61917
DOI: https://dx.doi.org/10.1109/TBCAS.2018.2853670
ISSN: 1932-4545
Publisher: Institute of Electrical and Electronics Engineers
Start Page: 1088
End Page: 1099
Journal / Book Title: IEEE Transactions on Biomedical Circuits and Systems
Volume: 12
Issue: 5
Copyright Statement: © 2018 The Author(s). This work is licensed under a Creative Commons Attribution 3.0 License. For more information, see http://creativecommons.org/licenses/by/3.0/
Sponsor/Funder: Engineering & Physical Science Research Council (EPSRC)
Funder's Grant Number: EP/M020975/1
Keywords: Science & Technology
Technology
Engineering, Biomedical
Engineering, Electrical & Electronic
Engineering
Chip-scale coil
implantable neural microsystem
integrated coil
mm-sized coil
microfabricated coil
near-field coupling
wirewound coil
wireless power transmission
SPIRAL INDUCTORS
WIRELESS POWER
RF ICS
TRANSMISSION
EXPRESSIONS
SILICON
MODEL
Animals
Equipment and Supplies
Prostheses and Implants
Ribs
Sheep
Transistors, Electronic
Wireless Technology
Ribs
Animals
Sheep
Equipment and Supplies
Prostheses and Implants
Transistors, Electronic
Wireless Technology
0903 Biomedical Engineering
0906 Electrical and Electronic Engineering
Electrical & Electronic Engineering
Publication Status: Published
Online Publication Date: 2018-07-20
Appears in Collections:Electrical and Electronic Engineering
Faculty of Engineering