In silico and in vivo investigations using an endocavitary photoplethysmography sensor for tissue viability monitoring
File(s)027001_1.pdf (2.98 MB)
Published version
Author(s)
Chatterjee, Subhasri
Patel, Zaibaa
Thaha, Mohamed A
Kyriacou, Panayiotis A
Type
Journal Article
Abstract
Significance: Colorectal cancer is one of the major causes of cancer-related deaths worldwide. Surgical removal of the cancerous growth is the primary treatment for this disease. A colorectal cancer surgery, however, is often unsuccessful due to the anastomotic failure that may occur following the surgical incision. Prevention of an anastomotic failure requires continuous monitoring of intestinal tissue viability during and after colorectal surgery. To date, no clinical technology exists for the dynamic and continuous monitoring of the intestinal perfusion.
Aim: A dual-wavelength indwelling bowel photoplethysmography (PPG) sensor for the continuous monitoring of intestinal viability was proposed and characterized through a set of in silico and in vivo investigations.
Approach: The in silico investigation was based on a Monte Carlo model that was executed to quantify the variables such as penetration depth and detected intensity with respect to the sensor–tissue separations and tissue perfusion. Utilizing the simulated information, an indwelling reflectance PPG sensor was designed and tested on 20 healthy volunteers. Two sets of in vivo studies were performed using the driving current intensities 20 and 40 mA for a comparative analysis, using buccal tissue as a proxy tissue-site.
Results: Both simulated and experimental results showed the efficacy of the sensor to acquire good signals through the “contact” to a “noncontact” separation of 5 mm. A very slow wavelength-dependent variation was shown in the detected intensity at the normal and hypoxic states of the tissue, whereas a decay in the intensity was found with the increasing submucosal-blood volume. The simulated detected-to-incident-photon-ratio and the experimental signal-to-noise ratio exhibited strong positive correlations, with the Pearson product-moment correlation coefficient R ranging between 0.65 and 0.87.
Conclusions: The detailed feasibility analysis presented will lead to clinical trials utilizing the proposed sensor.
Aim: A dual-wavelength indwelling bowel photoplethysmography (PPG) sensor for the continuous monitoring of intestinal viability was proposed and characterized through a set of in silico and in vivo investigations.
Approach: The in silico investigation was based on a Monte Carlo model that was executed to quantify the variables such as penetration depth and detected intensity with respect to the sensor–tissue separations and tissue perfusion. Utilizing the simulated information, an indwelling reflectance PPG sensor was designed and tested on 20 healthy volunteers. Two sets of in vivo studies were performed using the driving current intensities 20 and 40 mA for a comparative analysis, using buccal tissue as a proxy tissue-site.
Results: Both simulated and experimental results showed the efficacy of the sensor to acquire good signals through the “contact” to a “noncontact” separation of 5 mm. A very slow wavelength-dependent variation was shown in the detected intensity at the normal and hypoxic states of the tissue, whereas a decay in the intensity was found with the increasing submucosal-blood volume. The simulated detected-to-incident-photon-ratio and the experimental signal-to-noise ratio exhibited strong positive correlations, with the Pearson product-moment correlation coefficient R ranging between 0.65 and 0.87.
Conclusions: The detailed feasibility analysis presented will lead to clinical trials utilizing the proposed sensor.
Date Issued
2020-02-28
Date Acceptance
2020-02-18
Citation
Journal of Biomedical Optics, 2020, 25 (2), pp.1-16
ISSN
1083-3668
Publisher
Society of Photo-optical Instrumentation Engineers
Start Page
1
End Page
16
Journal / Book Title
Journal of Biomedical Optics
Volume
25
Issue
2
Copyright Statement
© The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.
Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. [DOI: 10.1117/1.JBO.25.2.027001]
Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. [DOI: 10.1117/1.JBO.25.2.027001]
License URL
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000590142200008&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Physical Sciences
Biochemical Research Methods
Optics
Radiology, Nuclear Medicine & Medical Imaging
Biochemistry & Molecular Biology
photoplethysmography
Monte Carlo
endocavitary sensor
signal-to-noise ratio
penetration depth
OPTICAL-PROPERTIES
PERFUSION
Publication Status
Published
Article Number
ARTN 027001
Date Publish Online
2020-02-28