Pre-clinical evaluation of a smartphone-assisted navigation system for surgical pin positioning in total knee arthroplasty
File(s) Bull+et+al_2026_Med._Eng._Phys._10.1088_1873-4030_aeb26c.pdf (508.3 KB)
Published online version
Author(s)
Type
Journal Article
Abstract
Total knee arthroplasty (TKA) success is highly dependent on accurate implant alignment, however, 80% of procedures globally are performed manually. Manual procedures have higher alignment errors compared to robotic platforms and patient-specific instrumentation, yet these have limitations of high cost and intraoperative inflexibility, respectively. This study aimed to evaluate the intraoperative pin positioning accuracy and flexibility of a smartphone-assisted navigation system using CT-derived surgical plans. Three participants used the navigation system to perform 108 guide pin insertions (54 femoral, 54 tibial) on surrogate bone models based on CT-derived surgical plans. Testing included both default mechanical axis targets (0°) and user-selected navigated targets ranging from -3° to 3°. Accuracy was assessed by comparing achieved pin trajectories with the preoperative surgical plan. Median absolute angular deviations achieved ranged from 0.9° to 1.6° across all component–plane combinations, with an overall median deviation of 1.3° (IQR 0.6–2.1°). The proportion of insertions within a predefined ±3° clinical accuracy threshold ranged from 83.3% to 100%, and all component–plane combinations demonstrated deviations significantly within this threshold (p < 0.00001). These findings support the technical feasibility of the smartphone-assisted navigation system for surgical pin positioning under controlled pre-clinical conditions. Further studies will evaluate the navigation system in cadaveric models with experienced orthopaedic surgeon participants.
Date Issued
2026-10-09
Date Acceptance
2026-10-09
Citation
Medical Engineering and Physics, 2026
ISSN
1350-4533
Publisher
IOP Publishing
Journal / Book Title
Medical Engineering and Physics
Copyright Statement
© 2026 The Author(s). Published on behalf of Institute of Physics and Engineering in Medicine by IOP Publishing Ltd.. As the Version of Record of this article is going to be / has been published on a gold open access basis under a CC BY 4.0 licence, this Accepted Manuscript is available for reuse under a CC BY 4.0 licence immediately.
License URL
Publication Status
Published online
Date Publish Online
2026-10-09