Within-patient comparison of His-bundle pacing, right ventricular pacing, and right ventricular pacing avoidance algorithms in patients with PR prolongation: Acute hemodynamic study
File(s)jce.14763.pdf (1.55 MB)
Published version
Author(s)
Type
Journal Article
Abstract
Aims
A prolonged PR interval may adversely affect ventricular filling and, therefore, cardiac function. AV delay can be corrected using right ventricular pacing (RVP), but this induces ventricular dyssynchrony, itself harmful. Therefore, in intermittent heart block, pacing avoidance algorithms are often implemented. We tested His‐bundle pacing (HBP) as an alternative.
Methods
Outpatients with a long PR interval (>200 ms) and intermittent need for ventricular pacing were recruited. We measured within‐patient differences in high‐precision hemodynamics between AV‐optimized RVP and HBP, as well as a pacing avoidance algorithm (Managed Ventricular Pacing [MVP]).
Results
We recruited 18 patients. Mean left ventricular ejection fraction was 44.3 ± 9%. Mean intrinsic PR interval was 266 ± 42 ms and QRS duration was 123 ± 29 ms. RVP lengthened QRS duration (+54 ms, 95% CI 42–67 ms, p < .0001) while HBP delivered a shorter QRS duration than RVP (−56 ms, 95% CI −67 to −46 ms, p < .0001). HBP did not increase QRS duration (−2 ms, 95% CI −8 to 13 ms, p = .6). HBP improved acute systolic blood pressure by mean of 5.0 mmHg (95% CI 2.8–7.1 mmHg, p < .0001) compared to RVP and by 3.5 mmHg (95% CI 1.9–5.0 mmHg, p = .0002) compared to the pacing avoidance algorithm. There was no significant difference in hemodynamics between RVP and ventricular pacing avoidance (p = .055).
Conclusions
HBP provides better acute cardiac function than pacing avoidance algorithms and RVP, in patients with prolonged PR intervals. HBP allows normalization of prolonged AV delays (unlike pacing avoidance) and does not cause ventricular dyssynchrony (unlike RVP). Clinical trials may be justified to assess whether these acute improvements translate into longer term clinical benefits in patients with bradycardia indications for pacing.
A prolonged PR interval may adversely affect ventricular filling and, therefore, cardiac function. AV delay can be corrected using right ventricular pacing (RVP), but this induces ventricular dyssynchrony, itself harmful. Therefore, in intermittent heart block, pacing avoidance algorithms are often implemented. We tested His‐bundle pacing (HBP) as an alternative.
Methods
Outpatients with a long PR interval (>200 ms) and intermittent need for ventricular pacing were recruited. We measured within‐patient differences in high‐precision hemodynamics between AV‐optimized RVP and HBP, as well as a pacing avoidance algorithm (Managed Ventricular Pacing [MVP]).
Results
We recruited 18 patients. Mean left ventricular ejection fraction was 44.3 ± 9%. Mean intrinsic PR interval was 266 ± 42 ms and QRS duration was 123 ± 29 ms. RVP lengthened QRS duration (+54 ms, 95% CI 42–67 ms, p < .0001) while HBP delivered a shorter QRS duration than RVP (−56 ms, 95% CI −67 to −46 ms, p < .0001). HBP did not increase QRS duration (−2 ms, 95% CI −8 to 13 ms, p = .6). HBP improved acute systolic blood pressure by mean of 5.0 mmHg (95% CI 2.8–7.1 mmHg, p < .0001) compared to RVP and by 3.5 mmHg (95% CI 1.9–5.0 mmHg, p = .0002) compared to the pacing avoidance algorithm. There was no significant difference in hemodynamics between RVP and ventricular pacing avoidance (p = .055).
Conclusions
HBP provides better acute cardiac function than pacing avoidance algorithms and RVP, in patients with prolonged PR intervals. HBP allows normalization of prolonged AV delays (unlike pacing avoidance) and does not cause ventricular dyssynchrony (unlike RVP). Clinical trials may be justified to assess whether these acute improvements translate into longer term clinical benefits in patients with bradycardia indications for pacing.
Date Issued
2020-10-05
Date Acceptance
2020-09-15
Citation
Journal of Cardiovascular Electrophysiology, 2020, 31 (11), pp.2964-2974
ISSN
1045-3873
Publisher
Wiley
Start Page
2964
End Page
2974
Journal / Book Title
Journal of Cardiovascular Electrophysiology
Volume
31
Issue
11
Copyright Statement
© 2020 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
License URL
Sponsor
British Heart Foundation
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000574941800001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Grant Number
RG/16/3/32175
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Cardiovascular System & Cardiology
AV optimization
His-bundle pacing
pacing avoidance algorithms
prolonged PR interval
DUAL-CHAMBER
CARDIAC-RESYNCHRONIZATION
HEART-FAILURE
INTERVAL
OUTCOMES
DYSFUNCTION
IMPROVEMENT
MORTALITY
BLOCK
Publication Status
Published
Date Publish Online
2020-09-25