Introduction
Clinical evaluation of deep learning (DL) tools is essential to compliment technical accuracy metrics. This study assessed the image quality of standard fetal head planes automatically-extracted from three-dimensional (3D) ultrasound fetal head volumes using a customised DL-algorithm.
Methods
Two observers retrospectively reviewed standard fetal head planes against pre-defined image quality criteria. Forty-eight images (29 transventricular, 19 transcerebellar) were selected from 91 transabdominal fetal scans (mean gestational age = 26 completed weeks, range = 20+5–32+3 weeks). Each had two-dimensional (2D) manually-acquired (2D-MA), 3D operator-selected (3D-OS) and 3D-DL automatically-acquired (3D-DL) images. The proportion of adequate images from each plane and modality, and the number of inadequate images per plane was compared for each method. Inter and intra-observer agreement of overall image quality was calculated.
Results
Sixty-seven percent of 3D-OS and 3D-DL transventricular planes were adequate quality. Forty-five percent of 3D-OS and 55% of 3D-DL transcerebellar planes were adequate.
Seventy-one percent of 3D-OS and 86% of 3D-DL transventricular planes failed with poor visualisation of intra-cranial structures. Eighty-six percent of 3D-OS and 80% of 3D-DL transcerebellar planes failed due to inadequate visualisation of cerebellar hemispheres. Image quality was significantly different between 2D and 3D, however, no significant difference between 3D-modalities was demonstrated (p < 0.005). Inter-observer agreement of transventricular plane adequacy was moderate for both 3D-modalities, and weak for transcerebellar planes.
Conclusion
The 3D-DL algorithm can automatically extract standard fetal head planes from 3D-head volumes of comparable quality to operator-selected planes. Image quality in 3D is inferior to corresponding 2D planes, likely due to limitations with 3D-technology and acquisition technique.
Implications for practice
Automated image extraction of standard planes from US-volumes could facilitate use of 3DUS in clinical practice, however image quality is dependent on the volume acquisition technique.
Clinical evaluation of deep learning (DL) tools is essential to compliment technical accuracy metrics. This study assessed the image quality of standard fetal head planes automatically-extracted from three-dimensional (3D) ultrasound fetal head volumes using a customised DL-algorithm.
Methods
Two observers retrospectively reviewed standard fetal head planes against pre-defined image quality criteria. Forty-eight images (29 transventricular, 19 transcerebellar) were selected from 91 transabdominal fetal scans (mean gestational age = 26 completed weeks, range = 20+5–32+3 weeks). Each had two-dimensional (2D) manually-acquired (2D-MA), 3D operator-selected (3D-OS) and 3D-DL automatically-acquired (3D-DL) images. The proportion of adequate images from each plane and modality, and the number of inadequate images per plane was compared for each method. Inter and intra-observer agreement of overall image quality was calculated.
Results
Sixty-seven percent of 3D-OS and 3D-DL transventricular planes were adequate quality. Forty-five percent of 3D-OS and 55% of 3D-DL transcerebellar planes were adequate.
Seventy-one percent of 3D-OS and 86% of 3D-DL transventricular planes failed with poor visualisation of intra-cranial structures. Eighty-six percent of 3D-OS and 80% of 3D-DL transcerebellar planes failed due to inadequate visualisation of cerebellar hemispheres. Image quality was significantly different between 2D and 3D, however, no significant difference between 3D-modalities was demonstrated (p < 0.005). Inter-observer agreement of transventricular plane adequacy was moderate for both 3D-modalities, and weak for transcerebellar planes.
Conclusion
The 3D-DL algorithm can automatically extract standard fetal head planes from 3D-head volumes of comparable quality to operator-selected planes. Image quality in 3D is inferior to corresponding 2D planes, likely due to limitations with 3D-technology and acquisition technique.
Implications for practice
Automated image extraction of standard planes from US-volumes could facilitate use of 3DUS in clinical practice, however image quality is dependent on the volume acquisition technique.
Date Issued
2021-05-01
Date Acceptance
2020-11-09
Citation
Radiography, 2021, 27 (2), pp.519-526
ISSN
1078-8174
Publisher
Elsevier
Start Page
519
End Page
526
Journal / Book Title
Radiography
Volume
27
Issue
2
Copyright Statement
© 2020 The College of Radiographers. Published by Elsevier Ltd. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
License URL
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000640791900038&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Radiology, Nuclear Medicine & Medical Imaging
Clinical evaluation
Fetal imaging
Quality assessment
Ultrasound
SONOGRAPHY
BRAIN
RELIABILITY
IMAGES
AUDIT
Publication Status
Published
Date Publish Online
2020-11-30