Brachial pressure waveforms: measuring and modelling
File(s)
Author(s)
Korolkova, Olga
Type
Thesis
Abstract
Hypertension is a major risk factor of stroke, myocardial infarction, heart failure and organ
damage. Although hypertension is currently defined based on the pressure measured in
the brachial artery, a number of studies have shown that central aortic blood pressure is
a better predictor of cardiovascular events. Central aortic pressure can only be measured
invasively, which is undesirable for a patient. It is preferable to measure central blood
pressure non-invasively with the same diagnostic efficacy since it is much simpler method
that does not require as much expertise.
Aortic pressure waveform, measured invasively in patients, is used as an input into nu-
merical one-dimensional model of the arterial network to generate brachial pressure wave-
forms. Modelled brachial pressure waveforms are compared with those measured invasively
(with a pressure wire) and non-invasively (with a cuff-based device). An unpredicted os-
cillatory behaviour of the pressure waveform in the brachial artery due to the occlusion of
the artery during the cuff inflation is observed.
Disagreement in systolic and diastolic pressures between invasive and non-invasive mea-
surements led to a study of the propagation of the pressure waveform from the brachial
artery through the soft tissue to the surface of the skin, where it is commonly measured. A
simplified upper arm model (consisting of one artery and a bone surrounded by soft tissue)
of the pressure propagation from the artery to the skin is built in the Comsol Multiphysics
software using poroelastic theory to describe the material properties. The brachial pres-
sure is approximated with a sinusoidal wave of varying frequency and used as an input in
the upper arm model. The results of this model show small uctuations in the cuff pressure. Since our hypothesis that the pressure waveform is being damped when propagating through the soft tissue has not been confirmed, some ideas for model improvement are also presented.
damage. Although hypertension is currently defined based on the pressure measured in
the brachial artery, a number of studies have shown that central aortic blood pressure is
a better predictor of cardiovascular events. Central aortic pressure can only be measured
invasively, which is undesirable for a patient. It is preferable to measure central blood
pressure non-invasively with the same diagnostic efficacy since it is much simpler method
that does not require as much expertise.
Aortic pressure waveform, measured invasively in patients, is used as an input into nu-
merical one-dimensional model of the arterial network to generate brachial pressure wave-
forms. Modelled brachial pressure waveforms are compared with those measured invasively
(with a pressure wire) and non-invasively (with a cuff-based device). An unpredicted os-
cillatory behaviour of the pressure waveform in the brachial artery due to the occlusion of
the artery during the cuff inflation is observed.
Disagreement in systolic and diastolic pressures between invasive and non-invasive mea-
surements led to a study of the propagation of the pressure waveform from the brachial
artery through the soft tissue to the surface of the skin, where it is commonly measured. A
simplified upper arm model (consisting of one artery and a bone surrounded by soft tissue)
of the pressure propagation from the artery to the skin is built in the Comsol Multiphysics
software using poroelastic theory to describe the material properties. The brachial pres-
sure is approximated with a sinusoidal wave of varying frequency and used as an input in
the upper arm model. The results of this model show small uctuations in the cuff pressure. Since our hypothesis that the pressure waveform is being damped when propagating through the soft tissue has not been confirmed, some ideas for model improvement are also presented.
Version
Open access
Date Issued
2013-06
Date Awarded
2013-07
Copyright Statement
Creative Commons Attribution NonCommercial NoDerivatives Licence
Advisor
Siggers, Jennifer
Parker, Kim
Hughes, Alun
Davies, Justin
Sponsor
Foundation for Circulatory Health
Publisher Department
Department of Bioengineering
Publisher Institution
Imperial College London
Qualification Level
Doctoral
Qualification Name
Doctor of Philosophy (PhD)