Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
File(s)
Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements.pdf (1.22 MB)
Supporting information
Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements.pdf (1.22 MB)
Published version
Author(s)
Gatehouse, Peter D
Rolf, Marijn P
Graves, Martin J
Hofman, Mark BM
Totman, John
more
Type
Journal Article
Abstract
Aims

Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems.
Methods and Results

In a multi-centre, multi-vendor study, breath-hold through-plane retrospectively ECG-gated phase contrast acquisitions, as are used clinically for aortic and pulmonary flow measurement, were applied to static gelatin phantoms in twelve 1.5 T CMR systems, using a velocity encoding range of 150 cm/s. No post-processing corrections of offsets were implemented. The greatest uncorrected velocity offset, taken as an average over a 'great vessel' region (30 mm diameter) located up to 70 mm in-plane distance from the magnet isocenter, ranged from 0.4 cm/s to 4.9 cm/s. It averaged 2.7 cm/s over all the planes and systems. By theoretical calculation, a velocity offset error of 0.6 cm/s (representing just 0.4% of a 150 cm/s velocity encoding range) is barely acceptable, potentially causing about 5% miscalculation of cardiac output and up to 10% error in shunt measurement.
Conclusion

In the absence of hardware or software upgrades able to reduce phase offset errors, all the systems tested appeared to require post-acquisition correction to achieve consistently reliable breath-hold measurements of flow. The effectiveness of offset correction software will still need testing with respect to clinical flow acquisitions.
Date Issued
2010-01-14
Date Acceptance
2010-01-14
Citation
Journal of Cardiovascular Magnetic Resonance, 2010, 12 (1)
URI
http://hdl.handle.net/10044/1/66215
DOI
https://www.dx.doi.org/10.1186/1532-429X-12-5
ISSN
1097-6647
Publisher
BioMed Central
Journal / Book Title
Journal of Cardiovascular Magnetic Resonance
Volume
12
Issue
1
Copyright Statement
© 2010 Gatehouse et al; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000274785200001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Radiology, Nuclear Medicine & Medical Imaging
Cardiovascular System & Cardiology
PULMONARY REGURGITATION
MITRAL REGURGITATION
BLOOD-FLOW
QUANTIFICATION
SEVERITY
MR
Publication Status
Published
Article Number
5
Date Publish Online
2010-01-14
About
Spiral Depositing with Spiral Publishing with Spiral Symplectic
Contact us
Open access team Report an issue
Other Services
Scholarly Communications Library Services
logo

Imperial College London

South Kensington Campus

London SW7 2AZ, UK

tel: +44 (0)20 7589 5111

Accessibility Modern slavery statement Cookie Policy

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback