A high spatio-temporal resolution optical pyrometer at the ORION laser facility
File(s)Orion_Pyrometry_2016.pdf (1.05 MB)
Published version
Author(s)
Type
Journal Article
Abstract
A streaked pyrometer has been designed to measure the temperature of ≈100 µm diameter heated
targets in the warm dense matter region. The diagnostic has picosecond time resolution. Spatial
resolution is limited by the streak camera to 4 µm in one dimension; the imaging system has
superior resolution of 1 µm. High light collection efficiency means that the diagnostic can transmit
a measurable quantity of thermal emission at temperatures as low as 1 eV to the detector. This is
achieved through the use of an f/1.4 objective, and a minimum number of reflecting and refracting
surfaces to relay the image over 8 m with no vignetting over a 0.4 mm field of view with
12.5× magnification. All the system optics are highly corrected, to allow imaging with minimal
aberrations over a broad spectral range. The detector is a highly sensitive Axis Photonique streak
camera with a P820PSU streak tube. For the first time, two of these cameras have been absolutely
calibrated at 1 ns and 2 ns sweep speeds under full operational conditions and over 8 spectral bands
between 425 nm and 650 nm using a high-stability picosecond white light source. Over this range the
cameras had a response which varied between 47 ± 8 and 14 ± 4 photons/count. The calibration of
the optical imaging system makes absolute temperature measurements possible. Color temperature
measurements are also possible due to the wide spectral range over which the system is calibrated;
two different spectral bands can be imaged onto different parts of the photocathode of the same streak
camera.
targets in the warm dense matter region. The diagnostic has picosecond time resolution. Spatial
resolution is limited by the streak camera to 4 µm in one dimension; the imaging system has
superior resolution of 1 µm. High light collection efficiency means that the diagnostic can transmit
a measurable quantity of thermal emission at temperatures as low as 1 eV to the detector. This is
achieved through the use of an f/1.4 objective, and a minimum number of reflecting and refracting
surfaces to relay the image over 8 m with no vignetting over a 0.4 mm field of view with
12.5× magnification. All the system optics are highly corrected, to allow imaging with minimal
aberrations over a broad spectral range. The detector is a highly sensitive Axis Photonique streak
camera with a P820PSU streak tube. For the first time, two of these cameras have been absolutely
calibrated at 1 ns and 2 ns sweep speeds under full operational conditions and over 8 spectral bands
between 425 nm and 650 nm using a high-stability picosecond white light source. Over this range the
cameras had a response which varied between 47 ± 8 and 14 ± 4 photons/count. The calibration of
the optical imaging system makes absolute temperature measurements possible. Color temperature
measurements are also possible due to the wide spectral range over which the system is calibrated;
two different spectral bands can be imaged onto different parts of the photocathode of the same streak
camera.
Date Issued
2016-09-13
Date Acceptance
2016-07-30
Citation
Review of Scientific Instruments, 2016, 87 (11)
ISSN
0034-6748
Publisher
AIP Publishing
Journal / Book Title
Review of Scientific Instruments
Volume
87
Issue
11
Copyright Statement
© 2016 American Institute of Physics. Review of Scientific Instruments 87, 11E546 (2016);
Sponsor
Engineering & Physical Science Research Council (EPSRC)
Engineering & Physical Science Research Council (E
Grant Number
EP/H500227/1
EP/R511547/1
Subjects
Science & Technology
Technology
Physical Sciences
Instruments & Instrumentation
Physics, Applied
Physics
Publication Status
Published
Article Number
11E546