A rare case of suture material obstructing the closure mechanism of a prosthetic aortic valve: a case report
Author(s)
Type
Journal Article
Abstract
Prosthetic aortic valve dysfunction presenting as aortic regurgitation is a complication of mechanical valve replacement. We describe a case of late valve dysfunction caused by an annular suture of excessive length obstructing the closure mechanism of a bileaflet prosthetic valve.We present this rare cause of valve dysfunction in an 80-year-old male patient who presented with haemolysis and dyspnoea. At the time of operation it was found that a long vertically positioned annular valve suture was interfering with the normal closure mechanism of one of the prosthetic leaflets causing eccentric regurgitation jets. These findings were misdiagnosed as paravalvular leaks on the preoperative transoesophageal echo. No paravalvular leak was identified intraoperatively. After removal of the responsible suture normal prosthetic valve function was restored.Whilst early aortic valve dysfunction caused by suture material has previously been reported, this is the first report of suture material causing late dysfunction.
Date Issued
2009-12-02
Date Acceptance
2009-12-02
Citation
Cases Journal, 2009, 2
ISSN
1757-1626
Publisher
BioMed Central
Journal / Book Title
Cases Journal
Volume
2
Copyright Statement
© 2009 The Authors; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0),
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Identifier
http://www.ncbi.nlm.nih.gov/pubmed/20062703
Subjects
1199 Other Medical And Health Sciences
Publication Status
Published
Coverage Spatial
England
Article Number
9126