Abstract
Purpose: To achieve rapid, low SAR super-resolution imaging by exploiting the characteristic magnetization off-resonance profile in SSFP.
Theory and Methods: In the presented technique, low flip angle unbalanced SSFP imaging is used to acquire a series of images at a low nominal resolution which are then combined in a super-resolution strategy analogous to non-linear structured illumination microscopy. This is demonstrated in principle via Bloch simulations and synthetic phantoms, and the performance is quantified in terms of point-spread function (PSF) and signal-to-noise ratio (SNR) for gray and white matter from field strengths of 0.35T to 9.4T. A k-space reconstruction approach is proposed to account for B0 effects. This was applied to reconstruct super-resolution images from a test object at 9.4T.
Results: Artifact-free super-resolution images were produced after incorporating sufficient preparation time for the magnetization to approach the steady state. High-resolution images of a test object were obtained at 9.4T, in the presence of considerable B0 inhomogeneity. For gray matter, the highest achievable resolution ranges from 3% of the acquired voxel dimension at 0.35T, to 9% at 9.4T. For white matter, this corresponds to 3% and 10% respectively. Compared to an equivalent segmented gradient echo acquisition at the optimal flip angle, with a fixed TR of 8ms, gray matter has up to 34% of the SNR at 9.4T while using a x10 smaller flip angle. For white matter, this corresponds to 29% with a x11 smaller flip angle.
Conclusion: This approach achieves high degrees of super-resolution enhancement with minimal RF power requirements.
Date Issued
2021-05-01
Date Acceptance
2020-10-20
Citation
Magnetic Resonance in Medicine, 2021, 85 (5), pp.2477-2489
URI
http://hdl.handle.net/10044/1/84896
URL
https://onlinelibrary.wiley.com/doi/10.1002/mrm.28593
DOI
https://www.dx.doi.org/10.1002/mrm.28593
ISSN
0740-3194
Publisher
Wiley
Start Page
2477
End Page
2489
Journal / Book Title
Magnetic Resonance in Medicine
Volume
85
Issue
5
Copyright Statement
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
License URL
https://creativecommons.org/licenses/by/4.0/
Sponsor
Wellcome Trust
UK DRI Ltd
UK DRI Ltd
UK DRI Ltd
Identifier
https://onlinelibrary.wiley.com/doi/10.1002/mrm.28593
Grant Number
220473/Z/20/Z
N/A
4050641385
N/A
Subjects
Magnetic Resonance Imaging
Image Processing, Computer-Assisted
Super-Resolution
High Resolution Imaging
Publication Status
Published
Date Publish Online
2020-11-17
About
Spiral Depositing with Spiral Publishing with Spiral Symplectic
Contact us
Open access team Report an issue
Other Services
Scholarly Communications Library Services
logo

Imperial College London

South Kensington Campus

London SW7 2AZ, UK

tel: +44 (0)20 7589 5111

Accessibility Modern slavery statement Cookie Policy

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback