High temporal resolution delayed analysis of clinical microdialysate streams
File(s)c7an01209h.pdf (1.95 MB)
Published version
Author(s)
Type
Journal Article
Abstract
This paper presents the use of tubing to store clinical microdialysis samples for delayed analysis with high temporal resolution, offering an alternative to traditional discrete offline microdialysis sampling. Samples stored in this way were found to be stable for up to 72 days at −80 °C. Examples of how this methodology can be applied to glucose and lactate measurement in a wide range of in vivo monitoring experiments are presented. This paper presents a general model, which allows for an informed choice of tubing parameters for a given storage time and flow rate avoiding high back pressure, which would otherwise cause the microdialysis probe to leak, while maximising temporal resolution.
Date Issued
2018-01-16
Date Acceptance
2017-10-19
Citation
Analyst, 2018, 143, pp.715-724
ISSN
1364-5528
Publisher
Royal Society of Chemistry
Start Page
715
End Page
724
Journal / Book Title
Analyst
Volume
143
Copyright Statement
This article is licensed under a Creative Commons Attribution 3.0 Unported Licence.
License URL
Sponsor
Engineering & Physical Science Research Council (EPSRC)
Wellcome Trust
Ovarian Cancer Action
Grant Number
EP/H009744/1
094912/Z/10/Z
N/A
Subjects
0301 Analytical Chemistry
0399 Other Chemical Sciences
Analytical Chemistry
Publication Status
Published