Device design modifications informed by in vitro testing of bacterial attachment reduce infection rates of cochlear implants in clinical practice
File(s) microorganisms-09-01809-v2.pdf (1.92 MB)
Published version
OA Location
Author(s)
Type
Journal Article
Abstract
Recalcitrant chronic infections of implanted medical devices are often linked to the presence of biofilms. The prevention and treatment of medical device-associated infections is a major source of antibiotic use and driver of antimicrobial resistance globally. Lowering the incidence of infection in patients that receive implanted medical devices could therefore significantly improve antibiotic stewardship and reduce patient morbidity. Here we determined if modifying the design of an implantable medical device to reduce bacterial attachment, impacted the incidence of device-associated infections in clinical practice. Since the 1980s cochlear implants have provided long-term treatment of sensorineural hearing deficiency in hundreds of thousands of patients world-wide. Nonetheless, a relatively small number of devices are surgically explanted each year due to unresolvable infections. Features associated with the accumulation of bacteria on the Cochlear™ Nucleus® CI24RE™ model of cochlear implant devices were identified using both in vitro bacterial attachment assays and examination of explanted devices. Macro-scale design modifications that reduced bacterial attachment in vitro were incorporated into the design of the CI500™ and Profile™ series of Nucleus implant. Analyses of mandatory post-market vigilance data of 198,757 CI24RE and 123,084 CI500/Profile series implantation surgeries revealed that these design modifications correlated with significantly reduced infection rates. This study demonstrates that a design-centric approach aimed at mitigating bacterial attachment was a simple, and effective means of reducing infections associated with Cochlear Nucleus devices. This approach is likely to be applicable to improving the designs of other implantable medical devices to reduce device-associated infections.
Date Issued
2021-08-25
Date Acceptance
2021-08-20
Citation
Microorganisms, 2021, 9 (9)
ISSN
2076-2607
Publisher
MDPI AG
Journal / Book Title
Microorganisms
Volume
9
Issue
9
Copyright Statement
© 2021 by the authors. Licensee MDPI, Basel, Switzerland.This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
License URL
Sponsor
Cystic Fibrosis Trust
Imperial College London
Grant Number
VIA 070
Publication Status
Published
Article Number
ARTN 1809