Does conventional intra-aortic balloon pump trigger timing produce optimal hemodynamic effects in vivo?
File(s)IJAO_2015_3_146-153.pdf (549.88 KB)
Published version
Author(s)
Kolyva, C
Pantalos, GM
Pepper, JR
Khir, AW
Type
Journal Article
Abstract
Purpose
The intra-aortic balloon pump (IABP) provides circulatory support through counterpulsation. The hemodynamic effects of the IABP may vary with assisting frequency and depend on IAB inflation/deflation timing. We aimed to assess in vivo the IABP benefits on coronary, aortic, and left ventricular hemodynamics at different assistance frequencies and trigger timings.
Methods
Six healthy, anesthetized, open-chest sheep received IABP support at 5 timing modes (EC, LC, CC, CE, CL, corresponding to early/late/conventional/conventional/conventional inflation and conventional/conventional/conventional/early/late deflation, respectively) with frequency 1:3 and 1:1. Aortic (Qao) and coronary (Qcor) flow, and aortic (Pao) and left ventricular (PLV) pressure were recorded simultaneously, with and without IABP support. Integrating systolic Qao yielded stroke volume (SV).
Results
EC at 1:1 produced the lowest end-diastolic Pao (59.5 ± 7.8 mmHg [EC], 63.4 ± 11.1 mmHg [CC]), CC at 1:1 the lowest systolic PLV (69.1 ± 6.5 mmHg [CC], 76.4 ± 6.5 mmHg [control]), CC at 1:1 the highest SV (88.5 ± 34.4 ml [CC], 76.6 ± 31.9 ml [control]) and CC at 1:3 the highest diastolic Qcor (187.2 ± 25.0 ml/min [CC], 149.9 ± 16.6 ml/min [control]). Diastolic Pao augmentation was enhanced by both assistance frequencies alike, and optimal timings were EC for 1:3 (10.4 ± 2.8 mmHg [EC], 6.7 ± 3.8 mmHg [CC]) and CC for 1:1 (10.8 ± 6.7 mmHg [CC], -3.0 ± 3.8 mmHg [control]).
Conclusions
In our experiments, neither a single frequency nor a single inflation/deflation timing, including conventional IAB timing, has shown superiority by uniformly benefiting all studied hemodynamic parameters. A choice of optimal frequency and IAB timing might need to be made based on individual patient hemodynamic needs rather than as a generalized protocol.
The intra-aortic balloon pump (IABP) provides circulatory support through counterpulsation. The hemodynamic effects of the IABP may vary with assisting frequency and depend on IAB inflation/deflation timing. We aimed to assess in vivo the IABP benefits on coronary, aortic, and left ventricular hemodynamics at different assistance frequencies and trigger timings.
Methods
Six healthy, anesthetized, open-chest sheep received IABP support at 5 timing modes (EC, LC, CC, CE, CL, corresponding to early/late/conventional/conventional/conventional inflation and conventional/conventional/conventional/early/late deflation, respectively) with frequency 1:3 and 1:1. Aortic (Qao) and coronary (Qcor) flow, and aortic (Pao) and left ventricular (PLV) pressure were recorded simultaneously, with and without IABP support. Integrating systolic Qao yielded stroke volume (SV).
Results
EC at 1:1 produced the lowest end-diastolic Pao (59.5 ± 7.8 mmHg [EC], 63.4 ± 11.1 mmHg [CC]), CC at 1:1 the lowest systolic PLV (69.1 ± 6.5 mmHg [CC], 76.4 ± 6.5 mmHg [control]), CC at 1:1 the highest SV (88.5 ± 34.4 ml [CC], 76.6 ± 31.9 ml [control]) and CC at 1:3 the highest diastolic Qcor (187.2 ± 25.0 ml/min [CC], 149.9 ± 16.6 ml/min [control]). Diastolic Pao augmentation was enhanced by both assistance frequencies alike, and optimal timings were EC for 1:3 (10.4 ± 2.8 mmHg [EC], 6.7 ± 3.8 mmHg [CC]) and CC for 1:1 (10.8 ± 6.7 mmHg [CC], -3.0 ± 3.8 mmHg [control]).
Conclusions
In our experiments, neither a single frequency nor a single inflation/deflation timing, including conventional IAB timing, has shown superiority by uniformly benefiting all studied hemodynamic parameters. A choice of optimal frequency and IAB timing might need to be made based on individual patient hemodynamic needs rather than as a generalized protocol.
Date Issued
2015-03-01
Date Acceptance
2015-01-14
Citation
International Journal of Artificial Organs, 2015, 38 (3), pp.146-153
ISSN
0391-3988
Publisher
Wichtig
Start Page
146
End Page
153
Journal / Book Title
International Journal of Artificial Organs
Volume
38
Issue
3
Copyright Statement
This article is available under CC BY-NC-ND license.
Subjects
Science & Technology
Technology
Life Sciences & Biomedicine
Engineering, Biomedical
Transplantation
Engineering
Intra-aortic balloon pump
Counterpulsation
Support frequency
Inflation timing
Deflation timing
COUNTERPULSATION
PERFORMANCE
PRESSURE
ARTERIES
SUPPORT
MODEL
FLOW
Animals
Female
Heart-Assist Devices
Hemodynamics
Intra-Aortic Balloon Pumping
Male
Sheep
Time Factors
Urology & Nephrology
0903 Biomedical Engineering
Publication Status
Published