Fully-modelled blood-focused variable inversion times for 3D late gadolinium-enhanced imaging
File(s) 1-s2.0-S0730725X22002296-main.pdf (3.37 MB)
Published version
Author(s)
Type
Journal Article
Abstract
Purpose
Variable heart rate during single-cycle inversion-recovery Late Gadolinium-Enhanced (LGE) scanning degrades image quality, which can be mitigated using Variable Inversion Times (VTIs) in real-time response to R-R interval changes. We investigate in vivo and in simulations an extension of a single-cycle VTI method previously applied in 3D LGE imaging, that now fully models the longitudinal magnetisation (fmVTI).
Methods
The VTI and fmVTI methods were used to perform 3D LGE scans for 28 3D LGE patients, with qualitative image quality scores assigned for left atrial wall clarity and total ghosting. Accompanying simulations of numerical phantom images were assessed in terms of ghosting of normal myocardium, blood, and myocardial scar.
Results
The numerical simulations for fmVTI showed a significant decrease in blood ghosting (VTI: 410 ± 710, fmVTI: 68 ± 40, p < 0.0005) and scar ghosting (VTI: 830 ± 1300, fmVTI: 510 ± 730, p < 0.02). Despite this, there was no significant change in qualitative image quality scores, either for left atrial wall clarity (VTI: 2.0 ± 1.0, fmVTI: 1.8 ± 1.0, p > 0.1) or for total ghosting (VTI: 1.9 ± 1.0, fmVTI: 2.0 ± 1.0, p > 0.7).
Conclusions
Simulations indicated reduced ghosting with the fmVTI method, due to reduced Mz variability in the blood signal. However, other sources of phase-encode ghosting and blurring appeared to dominate and obscure this finding in the patient studies available.
Variable heart rate during single-cycle inversion-recovery Late Gadolinium-Enhanced (LGE) scanning degrades image quality, which can be mitigated using Variable Inversion Times (VTIs) in real-time response to R-R interval changes. We investigate in vivo and in simulations an extension of a single-cycle VTI method previously applied in 3D LGE imaging, that now fully models the longitudinal magnetisation (fmVTI).
Methods
The VTI and fmVTI methods were used to perform 3D LGE scans for 28 3D LGE patients, with qualitative image quality scores assigned for left atrial wall clarity and total ghosting. Accompanying simulations of numerical phantom images were assessed in terms of ghosting of normal myocardium, blood, and myocardial scar.
Results
The numerical simulations for fmVTI showed a significant decrease in blood ghosting (VTI: 410 ± 710, fmVTI: 68 ± 40, p < 0.0005) and scar ghosting (VTI: 830 ± 1300, fmVTI: 510 ± 730, p < 0.02). Despite this, there was no significant change in qualitative image quality scores, either for left atrial wall clarity (VTI: 2.0 ± 1.0, fmVTI: 1.8 ± 1.0, p > 0.1) or for total ghosting (VTI: 1.9 ± 1.0, fmVTI: 2.0 ± 1.0, p > 0.7).
Conclusions
Simulations indicated reduced ghosting with the fmVTI method, due to reduced Mz variability in the blood signal. However, other sources of phase-encode ghosting and blurring appeared to dominate and obscure this finding in the patient studies available.
Date Issued
2023-05
Date Acceptance
2022-12-23
Citation
Magnetic Resonance Imaging, 2023, 98, pp.44-54
ISSN
0730-725X
Publisher
Elsevier
Start Page
44
End Page
54
Journal / Book Title
Magnetic Resonance Imaging
Volume
98
Copyright Statement
© 2023 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
License URL
Identifier
https://www.sciencedirect.com/science/article/pii/S0730725X22002296?via%3Dihub
Publication Status
Published
Date Publish Online
2022-12-26