A compact orthosis compliance monitoring device using pressure sensors and accelerometers: design and proof-of-concept testing
File(s) sensors-25-01352-v2.pdf (25.31 MB)
Published version
Author(s)
Devanand, Devi
Gardiner, Matthew
Kedgley, Angela
Type
Journal Article
Abstract
Monitoring orthosis compliance using patient diaries is subjective, as patients can overes-timate their levels of device use. An objective way to monitor compliance is required be-cause if an orthotic prescription is not followed the orthosis will not work as intended. This study aimed to develop and validate a device that monitors orthosis compliance ob-jectively using pressure and acceleration. Fifteen participants were recruited to test the de-vice’s ability to estimate wear time during the performance of several grip patterns and whilst completing selected activities of daily living. Sensor threshold values were used to discern whether users were wearing their orthosis or not. No differences between pressure sensor and accelerometer-based wear time estimations were found. The device’s pres-sure-based wear time estimations were found to have a specificity of 92.7 ± 16.4% and sen-sitivity of 74.0 ± 41.3%, whilst accelerometer-based wear time estimates had a specificity of 66.1 ± 34.7% and sensitivity of 86.2 ± 8.0%. This study successfully demonstrated the fea-sibility of monitoring hand orthosis compliance using pressure or acceleration. This de-vice has the potential to provide insight into the effectiveness of both existing and novel orthotics, benefitting both clinical practice and research.
Date Issued
2025-03-01
Date Acceptance
2025-02-19
Citation
Sensors, 2025, 25 (5)
ISSN
1424-8220
Publisher
MDPI
Journal / Book Title
Sensors
Volume
25
Issue
5
Copyright Statement
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/ licenses/by/4.0/).
License URL
Identifier
10.3390/s25051352
Subjects
Devanand, D.B.
Gardiner, M.D.
Kedgley, A.E. A Compact orthotics
design
hand
feasibility
compliance
adherence
pressure
sensor
accelerometer
wear time
Publication Status
Published
Article Number
1352