How to deliver personalized Cardiac Resynchronization Therapy through the precise measurement of the acute hemodynamic response: insights from the iSpot trial
File(s)
Author(s)
Type
Journal Article
Abstract
Introduction
New pacing technologies offer greater choice of left ventricular pacing sites and greater personalization of cardiac resynchronization therapy (CRT). The effects on cardiac function of novel pacing configurations are often compared using multi‐beat averages of acute hemodynamic measurements. In this analysis of the iSpot trial we explore whether this is sufficient.
Methods
The iSpot trial was an international, prospective, acute hemodynamic trial that assessed seven CRT configurations: Standard CRT, Multispot (posterolateral vein), and Multivein (anterior and posterior vein) pacing. Invasive and non‐invasive blood pressure, and LV dP/dtmax were recorded. Eight beats were recorded before and after an alternation from AAI to the tested pacing configuration and vice‐versa. Eight alternations were performed for each configuration at each of the 5 AV delays.
Results
25 patients underwent the full protocol of 8 alternations. Only 4 (16%) patients had a statistically significant >3mmHg improvement over conventional CRT configuration (posterolateral vein, distal electrode). However, if only one alternation was analyzed (standard multi‐beat averaging protocol), 15 (60%) patients falsely appeared to have a superior non‐conventional configuration. Responses to pacing were significantly correlated between the different hemodynamic measures: invasive SBP versus non‐invasive SBP r=0.82 (p<0.001); invasive SBP versus LV dP/dt r=0.57, r2=0.32 (p<0.001).
Conclusions
Current standard multi‐beat acquisition protocols are unfortunately unable to prevent false impressions of optimality arising in individual patients. Personalization processes need to include distinct repeated transitions to the tested pacing configuration in addition to averaging multiple beats. The need is not only during research stages, but also during clinical implementation.
New pacing technologies offer greater choice of left ventricular pacing sites and greater personalization of cardiac resynchronization therapy (CRT). The effects on cardiac function of novel pacing configurations are often compared using multi‐beat averages of acute hemodynamic measurements. In this analysis of the iSpot trial we explore whether this is sufficient.
Methods
The iSpot trial was an international, prospective, acute hemodynamic trial that assessed seven CRT configurations: Standard CRT, Multispot (posterolateral vein), and Multivein (anterior and posterior vein) pacing. Invasive and non‐invasive blood pressure, and LV dP/dtmax were recorded. Eight beats were recorded before and after an alternation from AAI to the tested pacing configuration and vice‐versa. Eight alternations were performed for each configuration at each of the 5 AV delays.
Results
25 patients underwent the full protocol of 8 alternations. Only 4 (16%) patients had a statistically significant >3mmHg improvement over conventional CRT configuration (posterolateral vein, distal electrode). However, if only one alternation was analyzed (standard multi‐beat averaging protocol), 15 (60%) patients falsely appeared to have a superior non‐conventional configuration. Responses to pacing were significantly correlated between the different hemodynamic measures: invasive SBP versus non‐invasive SBP r=0.82 (p<0.001); invasive SBP versus LV dP/dt r=0.57, r2=0.32 (p<0.001).
Conclusions
Current standard multi‐beat acquisition protocols are unfortunately unable to prevent false impressions of optimality arising in individual patients. Personalization processes need to include distinct repeated transitions to the tested pacing configuration in addition to averaging multiple beats. The need is not only during research stages, but also during clinical implementation.
Date Issued
2019-09
Date Acceptance
2019-05-19
Citation
Journal of Cardiovascular Electrophysiology, 2019, 30 (9), pp.1610-1619
ISSN
1045-3873
Publisher
Wiley
Start Page
1610
End Page
1619
Journal / Book Title
Journal of Cardiovascular Electrophysiology
Volume
30
Issue
9
Copyright Statement
© 2019 Owner. This is the accepted version of the following article: [Shun‐Shin, M. J., Miyazawa, A. A., Keene, D. , Sterliński, M. , Sokal, A. , Heuverswyn, F. , Rinaldi, C. A., Cornelussen, R. , Stegemann, B. , Francis, D. P. and Whinnett, Z. (2019), How to deliver personalized Cardiac Resynchronization Therapy through the precise measurement of the acute hemodynamic response: insights from the iSpot trial. J Cardiovasc Electrophysiol. Accepted Author Manuscript. doi:10.1111/jce.14001], which has been published in final form at https://doi.org/10.1111/jce.14001.
Sponsor
Foundation for Circulatory Health
British Heart Foundation
British Heart Foundation
British Heart Foundation
Identifier
https://onlinelibrary.wiley.com/doi/full/10.1111/jce.14001
Grant Number
ICCH/12/5039
FS/14/25/30676
FS/14/27/30752
CS/15/3/31405
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Cardiovascular System & Cardiology
cardiac resynchronization
heart failure
MultiSpot
MultiVein
optimization
INTERVENTRICULAR DELAY OPTIMIZATION
AV DELAY
HEART-FAILURE
PACING SITES
ALGORITHMS
IMPROVES
DESIGN
MultiSpot
MultiVein
cardiac resynchronization
heart failure
optimization
1102 Cardiorespiratory Medicine and Haematology
Cardiovascular System & Hematology
Publication Status
Published
Date Publish Online
2019-05-22