Myocardial first-pass perfusion imaging with hybrid-EPI: frequency-offsets and potential artefacts
Author(s)
Ferreira, PF
Gatehouse, PD
Firmin, DN
Type
Journal Article
Abstract
Background
First-pass myocardial perfusion is often imaged with a tailored hybrid centric interleaved echo-planar-imaging sequence, providing rapid image acquisition with good contrast enhancement. The centric interleaved phase-encode order minimises the effective time-of-echo but it is sensitive to frequency-offsets. This short article aims to show possible artefacts that might originate with this sequence, in the context of first-pass perfusion imaging, when frequency-offsets are present. Non-uniform magnitude modulation effects were also analysed.
Methods
Numerical and phantom simulations were used to illustrate the effects of frequency-offsets and non-uniform magnitude modulation with this sequence in a typical perfusion protocol. In vivo data was post-processed to analyse the h-EPI’s sensitivity to the frequency-offsets.
Results
The centric phase-order was shown to be highly sensitive to frequency-offsets due to its symmetrical phase slope. Resulting artefacts include blurring, and splitting of the image into two identical copies along the phase-encode direction. It was also shown that frequency-offsets can introduce signal loss and ghosting of the right ventricle signal into the myocardium. The in vivo results were confirmed by numerical and phantom simulations. Magnitude modulation effects were found to be small.
Conclusions
Imaging first-pass myocardial perfusion with an hybrid centric echo-planar-imaging sequence can be corrupted with ghosting and splitting of the image due to frequency-offsets.
First-pass myocardial perfusion is often imaged with a tailored hybrid centric interleaved echo-planar-imaging sequence, providing rapid image acquisition with good contrast enhancement. The centric interleaved phase-encode order minimises the effective time-of-echo but it is sensitive to frequency-offsets. This short article aims to show possible artefacts that might originate with this sequence, in the context of first-pass perfusion imaging, when frequency-offsets are present. Non-uniform magnitude modulation effects were also analysed.
Methods
Numerical and phantom simulations were used to illustrate the effects of frequency-offsets and non-uniform magnitude modulation with this sequence in a typical perfusion protocol. In vivo data was post-processed to analyse the h-EPI’s sensitivity to the frequency-offsets.
Results
The centric phase-order was shown to be highly sensitive to frequency-offsets due to its symmetrical phase slope. Resulting artefacts include blurring, and splitting of the image into two identical copies along the phase-encode direction. It was also shown that frequency-offsets can introduce signal loss and ghosting of the right ventricle signal into the myocardium. The in vivo results were confirmed by numerical and phantom simulations. Magnitude modulation effects were found to be small.
Conclusions
Imaging first-pass myocardial perfusion with an hybrid centric echo-planar-imaging sequence can be corrupted with ghosting and splitting of the image due to frequency-offsets.
Date Issued
2012-06-25
Date Acceptance
2012-06-17
Citation
Journal of Cardiovascular Magnetic Resonance, 2012, 14, pp.44-44
ISSN
1097-6647
Publisher
BioMed Central
Start Page
44
End Page
44
Journal / Book Title
Journal of Cardiovascular Magnetic Resonance
Volume
14
Copyright Statement
© 2012 Ferreira et al; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative
Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and
reproduction in any medium, provided the original work is properly cited.
Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and
reproduction in any medium, provided the original work is properly cited.
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Radiology, Nuclear Medicine & Medical Imaging
Cardiovascular System & Cardiology
CARDIAC & CARDIOVASCULAR SYSTEMS
RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
CORONARY-ARTERY-DISEASE
MAGNETIC-RESONANCE
BLOOD SIGNAL
MRI
SEQUENCES
Artifacts
Echo-Planar Imaging
Heart
Humans
Image Enhancement
Myocardial Perfusion Imaging
Myocardium
Phantoms, Imaging
Reproducibility of Results
1102 Cardiovascular Medicine And Haematology
Nuclear Medicine & Medical Imaging
Publication Status
Published
Article Number
1