Effect of temperature, wavelength, pH, ion pair reagents and organic modifi ers’ concentration on the elution of cystatin C. stability of mobile phase
Author(s)
Fayyad, Manar Khalid
Misha, Adel Khalil
Al-Musaimi, Othman Ibrahem Yousef
Type
Journal Article
Abstract
Robustness of an analytical chromatographic method for separation of cystatin c has been veri fi ed. Changes in many parameters were carried out, such as, wavelength, column oven, mobile phase composition, chromatographic column.
Imperative changes have altered the ef fi ciency of the chromatographic separation; such changes include pH alteration of the mobile phase as well as alkyl sulfonate molarity changing.
All robustness conditions showed no major effect on the chromatographic separation of the analyte except with the changes related to TFA and alkyl sulfonate ion pair reagents. Peak area RSD, asymmetry and No. of theoretical plates were < 0.7%, < 1.2 and > 10,000, respectively. Results obtained using mobile phase after 6 months of storage have proven its stability and possibility of use. Gradient elution mode was utilized to elute cystatin c with a UV detection of 224 nm. Ace and Waters C8 (150 x 4.6 mm i.d., 5 μ m) as chromatographic columns were used.
Imperative changes have altered the ef fi ciency of the chromatographic separation; such changes include pH alteration of the mobile phase as well as alkyl sulfonate molarity changing.
All robustness conditions showed no major effect on the chromatographic separation of the analyte except with the changes related to TFA and alkyl sulfonate ion pair reagents. Peak area RSD, asymmetry and No. of theoretical plates were < 0.7%, < 1.2 and > 10,000, respectively. Results obtained using mobile phase after 6 months of storage have proven its stability and possibility of use. Gradient elution mode was utilized to elute cystatin c with a UV detection of 224 nm. Ace and Waters C8 (150 x 4.6 mm i.d., 5 μ m) as chromatographic columns were used.
Date Issued
2010-10-01
Date Acceptance
2010-09-29
Citation
Journal of Analytical & Bioanalytical Techniques, 2010, 01 (01), pp.1-4
ISSN
2155-9872
Publisher
OMICS Publishing Group
Start Page
1
End Page
4
Journal / Book Title
Journal of Analytical & Bioanalytical Techniques
Volume
01
Issue
01
Copyright Statement
© 2010 Fayyad MK, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
License URL
Identifier
https://www.omicsonline.org/effect-of-temperature-wavelength-ph-ion-pair-reagents-and-organic-modifiers-concentration-on-the-elution-of-cystatin-c-stability-of-mobile-phase-2155-9872.1000103.php?aid=85
Publication Status
Published