Aim
Anorectal physiology tests provide a functional assessment of the anal canal. The aim of this study was to compare the results generated by standard high‐resolution water‐perfused manometry (WPM) with the newer THD® Anopress manometry system.
Method
This was a prospective observational study. Conventional manometry was carried out using a water‐perfused catheter with high‐resolution manometry and compared with the Anopress system with air‐filled catheters. All patients underwent the two procedures successively in a randomized order. Time to arrive at the resting pressure plateau, resting, squeeze, straining pressure and visual analogue scale (VAS) scores for pain were recorded. A qualitative analysis of the two devices was performed.
Results
Between 2016 and 2017, 60 patients were recruited. The time from insertion of the catheter to arriving at the resting pressure plateau was significantly lower with the Anopress compared with WPM: 12 s [interquartile range (IQR) 10–17 s] versus 100 s (IQR 67–121 s) (P < 0.001). A strong correlation between the manometric values of WPM and the Anopress was observed. Both procedures were well tolerated, although the VAS score for insertion of the WPM catheter was significantly higher. The Anopress was easier to use and more time‐efficient than the WPM.
Conclusion
The pressure values obtained with Anopress correlated well with those of conventional manometry. The Anopress has the advantage of being less time‐consuming, user‐friendly and better tolerated by patients.
Anorectal physiology tests provide a functional assessment of the anal canal. The aim of this study was to compare the results generated by standard high‐resolution water‐perfused manometry (WPM) with the newer THD® Anopress manometry system.
Method
This was a prospective observational study. Conventional manometry was carried out using a water‐perfused catheter with high‐resolution manometry and compared with the Anopress system with air‐filled catheters. All patients underwent the two procedures successively in a randomized order. Time to arrive at the resting pressure plateau, resting, squeeze, straining pressure and visual analogue scale (VAS) scores for pain were recorded. A qualitative analysis of the two devices was performed.
Results
Between 2016 and 2017, 60 patients were recruited. The time from insertion of the catheter to arriving at the resting pressure plateau was significantly lower with the Anopress compared with WPM: 12 s [interquartile range (IQR) 10–17 s] versus 100 s (IQR 67–121 s) (P < 0.001). A strong correlation between the manometric values of WPM and the Anopress was observed. Both procedures were well tolerated, although the VAS score for insertion of the WPM catheter was significantly higher. The Anopress was easier to use and more time‐efficient than the WPM.
Conclusion
The pressure values obtained with Anopress correlated well with those of conventional manometry. The Anopress has the advantage of being less time‐consuming, user‐friendly and better tolerated by patients.
Date Issued
2020-08-01
Date Acceptance
2020-01-22
Citation
Colorectal Disease, 2020, 22 (8), pp.923-930
ISSN
1462-8910
Publisher
Wiley
Start Page
923
End Page
930
Journal / Book Title
Colorectal Disease
Volume
22
Issue
8
Copyright Statement
© 2020 The Authors. Colorectal Disease published by John Wiley & Sons Ltd on behalf of Association of Coloproctology of Great Britain and Ireland.
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
License URL
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000555088900009&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Gastroenterology & Hepatology
Surgery
Anopress
anorectal manometry
anorectal physiology tests
faecal incontinence
high-resolution manometry
physiology
water-perfused manometry
Publication Status
Published
Date Publish Online
2020-01-28