Development and validation of ester impregnated pH strips for locating nasogastric feeding tubes in the stomach-a multicentre prospective diagnostic performance study.
File(s)Ni2021_Article_DevelopmentAndValidationOfEste.pdf (1.2 MB)
Published version
Author(s)
Type
Journal Article
Abstract
BACKGROUND: NG (nasogastric) tubes are used worldwide as a means to provide enteral nutrition. Testing the pH of tube aspirates prior to feeding is commonly used to verify tube location before feeding or medication. A pH at or lower than 5.5 was taken as evidence for stomach intubation. However, the existing standard pH strips lack sensitivity, especially in patients receiving feeding and antacids medication. We developed and validated a first-generation ester-impregnated pH strip test to improve the accuracy towards gastric placements in adult population receiving routine NG-tube feeding. The sensitivity was improved by its augmentation with the action of human gastric lipase (HGL), an enzyme specific to the stomach. METHODS: We carried out a multi-centred, prospective, two-gate diagnostic accuracy study on patients who require routine NG-tube feeding in 10 NHS hospitals comparing the sensitivity of the novel pH strip to the standard pH test, using either chest X-rays or, in its absence, clinical observation of the absence of adverse events as the reference standard. We also tested the novel pH strips in lung aspirates from patients undergoing oesophageal cancer surgeries using visual inspection as the reference standard. We simulated health economics using a decision analytic model and carried out adoption studies to understand its route to commercialisation. The primary end point is the sensitivity of novel and standard pH tests at the recommended pH cut-off of 5.5. RESULTS: A total of 6400 ester-impregnated pH strips were prepared based on an ISO13485 quality management system. A total of 376 gastric samples were collected from adult patients in 10 NHS hospitals who were receiving routine NG-tube feeding. The sensitivities of the standard and novel pH tests were respectively 49.2% (95% CI 44.1‑54.3%) and 70.2% (95% CI 65.6‑74.8%) under pH cut-off of 5.5 and the novel test has a lung specificity of 89.5% (95% CI 79.6%, 99.4%). Our simulation showed that using the novel test can potentially save 132 unnecessary chest X-rays per check per every 1000 eligible patients, or direct savings of £4034 to the NHS. CONCLUSIONS: The novel pH test correctly identified significantly more patients with tubes located inside the stomach compared to the standard pH test used widely by the NHS. TRIAL REGISTRATION: http://www.isrctn.com/ISRCTN11170249 , Registered 21 June 2017-retrospectively registered.
Date Issued
2021-12-14
Date Acceptance
2021-11-24
Citation
Diagnostic and Prognostic Research, 2021, 5 (1), pp.1-10
ISSN
2397-7523
Publisher
BioMed Central
Start Page
1
End Page
10
Journal / Book Title
Diagnostic and Prognostic Research
Volume
5
Issue
1
Copyright Statement
© The Author(s). 2021 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License,
which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give
appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if
changes were made. The images or other third party material in this article are included in the article's Creative Commons
licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons
licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain
permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give
appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if
changes were made. The images or other third party material in this article are included in the article's Creative Commons
licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons
licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain
permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
License URL
Sponsor
National Institute for Health Research
Identifier
https://www.ncbi.nlm.nih.gov/pubmed/34903303
PII: 10.1186/s41512-021-00111-9
Grant Number
RDB21 79560
Subjects
Diagnostics, Nutrition, NG-tubes, Health economics, Adoption barriers
Publication Status
Published
Coverage Spatial
England
Date Publish Online
2021-12-14