Abstract
We report results from an extensive set of measurements of the
β
-decay response in liquid xenon. These measurements are derived from high-statistics calibration data from injected sources of both
3
H
and
14
C
in the LUX detector. The mean light-to-charge ratio is reported for 13 electric field values ranging from 43 to
491
V
/
cm
, and for energies ranging from 1.5 to 145 keV.
β
-decay response in liquid xenon. These measurements are derived from high-statistics calibration data from injected sources of both
3
H
and
14
C
in the LUX detector. The mean light-to-charge ratio is reported for 13 electric field values ranging from 43 to
491
V
/
cm
, and for energies ranging from 1.5 to 145 keV.
Date Issued
2019-07-10
Date Acceptance
2019-07-01
Citation
Physical Review D: Particles, Fields, Gravitation and Cosmology, 2019, 100 (2)
ISSN
1550-2368
Publisher
American Physical Society
Journal / Book Title
Physical Review D: Particles, Fields, Gravitation and Cosmology
Volume
100
Issue
2
Copyright Statement
© 2019 American Physical Society. Cite this article as D. S. Akerib et al. (LUX Collaboration) Phys. Rev. D 100, 022002 DOI: https://doi.org/10.1103/PhysRevD.100.022002
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000474874900001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Physical Sciences
Astronomy & Astrophysics
Physics, Particles & Fields
Physics
Publication Status
Published
Article Number
022002
Date Publish Online
2019-07-10