Nickel hypersensitivity and endovascular devices: a systematic review and meta-analysis
File(s)Nickel_Heart Clean 2.docx (91.13 KB)
Accepted version
Author(s)
Gueroult, Aurelien
Al-Balah, A
Shalhoub, Joseph
Davies, Alun
Type
Journal Article
Abstract
Objective Nickel allergy is common; endovascular specialists are often confronted with nickel allergic patients ahead of the implantation of endovascular devices, many of which are nickel-containing. Our aim was to elucidate whether nickel hypersensitivity is significantly associated with worse or adverse outcomes after placement of a nickel-containing endovascular device.
Methods Inclusion criteria were: endovascular and transcatheter procedures for coronary, structural heart, neurovascular and peripheral vascular pathology involving nickel-allergic patients. All adverse outcomes were included as defined by included studies. A systematic review and meta-analysis were undertaken using a random-effects model. Searches of MEDLINE and EMBASE were conducted for articles published 1947–2019.
Results 190 records were identified, 78 articles were included for qualitative synthesis and 15 met criteria for meta-analysis. Patch-test confirmed nickel allergy was associated with an increased risk of adverse outcomes following implantation of a nickel-containing endovascular device (n=14 articles, 1740 patients; OR 2.61, 95% CI 1.41 to 4.85). This finding further was observed in coronary (n=12 articles, 1624 patients; OR 1.94, 95% CI 1.16 to 3.23) and structural heart subgroups (n=2 articles, 83 patients; OR 52.28, 95% CI 1.31 to 2079.14), but not in the neurovascular subgroup (n=1 article, 33 patients; OR 3.04, 95% CI 0.59 to 15.72) or with a patient-reported history of nickel allergy (n=2 articles, 207 patients; OR 2.14, 95% CI 0.23 to 19.70).
Conclusions Patch-tested nickel allergy is associated with an increased risk of adverse outcomes following endovascular device implantation and alternative treatment options should be considered. Specialists faced with patients’ self-reporting nickel allergy should consider proceeding to diagnostic patch-testing.
Methods Inclusion criteria were: endovascular and transcatheter procedures for coronary, structural heart, neurovascular and peripheral vascular pathology involving nickel-allergic patients. All adverse outcomes were included as defined by included studies. A systematic review and meta-analysis were undertaken using a random-effects model. Searches of MEDLINE and EMBASE were conducted for articles published 1947–2019.
Results 190 records were identified, 78 articles were included for qualitative synthesis and 15 met criteria for meta-analysis. Patch-test confirmed nickel allergy was associated with an increased risk of adverse outcomes following implantation of a nickel-containing endovascular device (n=14 articles, 1740 patients; OR 2.61, 95% CI 1.41 to 4.85). This finding further was observed in coronary (n=12 articles, 1624 patients; OR 1.94, 95% CI 1.16 to 3.23) and structural heart subgroups (n=2 articles, 83 patients; OR 52.28, 95% CI 1.31 to 2079.14), but not in the neurovascular subgroup (n=1 article, 33 patients; OR 3.04, 95% CI 0.59 to 15.72) or with a patient-reported history of nickel allergy (n=2 articles, 207 patients; OR 2.14, 95% CI 0.23 to 19.70).
Conclusions Patch-tested nickel allergy is associated with an increased risk of adverse outcomes following endovascular device implantation and alternative treatment options should be considered. Specialists faced with patients’ self-reporting nickel allergy should consider proceeding to diagnostic patch-testing.
Date Issued
2022-10-13
Date Acceptance
2021-09-23
Citation
Heart, 2022, 108 (21), pp.1707-1715
ISSN
1355-6037
Publisher
BMJ Publishing Group
Start Page
1707
End Page
1715
Journal / Book Title
Heart
Volume
108
Issue
21
Copyright Statement
© Author(s) (or their employer(s)) 2021. No commercial re-use. See rights and permissions. Published by BMJ. This article has been accepted for publication in Heart following peer review. The definitive copyedited, typeset version Guéroult AM, Al-Balah A, Davies AH, et alNickel hypersensitivity and endovascular devices: a systematic review and meta-analysisHeart Published Online First: 26 October 2021. doi: 10.1136/heartjnl-2021-319940
Identifier
https://heart.bmj.com/content/early/2021/10/26/heartjnl-2021-319940.info
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Cardiovascular System & Cardiology
aortic diseases
cardiac surgical procedures
coronary artery disease
endovascular procedures
heart defects
congenital
ATRIAL SEPTAL-DEFECT
METAL ALLERGY
STENT IMPLANTATION
PATIENT
DERMATITIS
LESIONS
GOLD
aortic diseases
cardiac surgical procedures
congenital
coronary artery disease
endovascular procedures
heart defects
Endovascular Procedures
Humans
Hypersensitivity
Nickel
Patch Tests
Treatment Outcome
Cardiovascular System & Hematology
1102 Cardiorespiratory Medicine and Haematology
1103 Clinical Sciences
Publication Status
Published
Date Publish Online
2021-10-26