Abstract
Background
Although novel hepatitis C (HCV) RNA point-of-care technology has the potential to enhance diagnosis in resource-limited settings, very little real-world validation of their utility exists. We evaluate the performance of HCV RNA quantification using the Xpert® HCV Viral Load Fingerstick assay (Xpert® HCV VL Fingerstick assay) as compared to the WHO pre-qualified plasma Xpert® HCV viral load assay among people who inject drugs (PWID) attending an opioid agonist therapy (OAT) clinic in Dar-es-Salaam, Tanzania.
Methods
Between December 2018 and February 2019 consecutive HCV seropositive PWID attending the OAT clinic provided paired venous and finger-stick samples for HCV RNA quantification. These were processed on-site using the GeneXpert® platform located at the Central tuberculosis reference laboratory.
Results
A total of 208 out of 220 anti-HCV positive participants recruited (94.5%) had a valid Xpert® HCV VL result available; 126 (61%; (95% CI 53.8-67.0) had detectable and quantifiable HCV RNA. 188 (85%) had paired plasma and finger-stick whole blood samples; the sensitivity and specificity for the quantification of HCV RNA levels were 99.1% and 98.7% respectively. There was an excellent correlation (R2=0.95) and concordance (mean difference 0.13 IU/mL, (95% CI -0.9 to 0.16 IU/mL) in HCV RNA levels between plasma samples and finger-stick samples.
Conclusion
This study found excellent performance of the Xpert® HCV VL Fingerstick assay for HCV RNA detection and quantification in an African-field setting. Its clinical utility represents an important watershed in overcoming existing challenges to HCV diagnosis, which should play a crucial role in HCV elimination in Africa.
Although novel hepatitis C (HCV) RNA point-of-care technology has the potential to enhance diagnosis in resource-limited settings, very little real-world validation of their utility exists. We evaluate the performance of HCV RNA quantification using the Xpert® HCV Viral Load Fingerstick assay (Xpert® HCV VL Fingerstick assay) as compared to the WHO pre-qualified plasma Xpert® HCV viral load assay among people who inject drugs (PWID) attending an opioid agonist therapy (OAT) clinic in Dar-es-Salaam, Tanzania.
Methods
Between December 2018 and February 2019 consecutive HCV seropositive PWID attending the OAT clinic provided paired venous and finger-stick samples for HCV RNA quantification. These were processed on-site using the GeneXpert® platform located at the Central tuberculosis reference laboratory.
Results
A total of 208 out of 220 anti-HCV positive participants recruited (94.5%) had a valid Xpert® HCV VL result available; 126 (61%; (95% CI 53.8-67.0) had detectable and quantifiable HCV RNA. 188 (85%) had paired plasma and finger-stick whole blood samples; the sensitivity and specificity for the quantification of HCV RNA levels were 99.1% and 98.7% respectively. There was an excellent correlation (R2=0.95) and concordance (mean difference 0.13 IU/mL, (95% CI -0.9 to 0.16 IU/mL) in HCV RNA levels between plasma samples and finger-stick samples.
Conclusion
This study found excellent performance of the Xpert® HCV VL Fingerstick assay for HCV RNA detection and quantification in an African-field setting. Its clinical utility represents an important watershed in overcoming existing challenges to HCV diagnosis, which should play a crucial role in HCV elimination in Africa.
Date Issued
2020-03
Date Acceptance
2019-11-15
Citation
Liver International, 2020, 40 (3), pp.514-521
ISSN
1478-3223
Publisher
Wiley
Start Page
514
End Page
521
Journal / Book Title
Liver International
Volume
40
Issue
3
Copyright Statement
© 2019 The Authors. Liver International published by John Wiley & Sons Ltd
This is an open access article under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
This is an open access article under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Sponsor
Wellcome Trust
Wellcome Trust
Wellcome Trust
Identifier
https://onlinelibrary.wiley.com/doi/full/10.1111/liv.14315
Grant Number
097816/Z/11/ZR
097816/Z/11/A
105603/Z/14/Z
Subjects
Science & Technology
Life Sciences & Biomedicine
Gastroenterology & Hepatology
hepatitis C viurs (HCV) diagnosis
people who inject drugs (PWID)
point-of-care (POC)
sub-Saharan Africa
Xpert
HEPATITIS-C
PERFORMANCE EVALUATION
RNA
ELIMINATION
CARE
PREVALENCE
DIAGNOSIS
LESSONS
ACCESS
IMPACT
Xpert
hepatitis C viurs (HCV) diagnosis
people who inject drugs (PWID)
point-of-care (POC)
sub-Saharan Africa
Gastroenterology & Hepatology
1103 Clinical Sciences
Publication Status
Published
Date Publish Online
2019-11-28