Clinical and laboratory evaluation of SARS-CoV-2 lateral flow assays for use in a national COVID-19 sero-prevalence survey
File(s)
Author(s)
Type
Journal Article
Abstract
Background
Accurate antibody tests are essential to monitor the SARS-CoV-2 pandemic. Lateral flow immunoassays (LFIAs) can deliver testing at scale. However, reported performance varies, and sensitivity analyses have generally been conducted on serum from hospitalised patients. For use in community testing, evaluation of finger-prick self-tests, in non-hospitalised individuals, is required.
Methods
Sensitivity analysis was conducted on 276 non-hospitalised participants. All had tested positive for SARS-CoV-2 by RT-PCR and were ≥21d from symptom-onset. In phase I we evaluated five LFIAs in clinic (with finger-prick) and laboratory (with blood and sera) in comparison to a) PCR-confirmed infection and b) presence of SARS-CoV-2 antibodies on two “in-house” ELISAs. Specificity analysis was performed on 500 pre-pandemic sera. In phase II, six additional LFIAs were assessed with serum.
Findings
95% (95%CI [92.2, 97.3]) of the infected cohort had detectable antibodies on at least one ELISA. LFIA sensitivity was variable, but significantly inferior to ELISA in 8/11 assessed. Of LFIAs assessed in both clinic and laboratory, finger-prick self-test sensitivity varied from 21%-92% vs PCR-confirmed cases and 22%-96% vs composite ELISA positives. Concordance between finger-prick and serum testing was at best moderate (kappa 0.56) and, at worst, slight (kappa 0.13). All LFIAs had high specificity (97.2% - 99.8%).
Interpretation
LFIA sensitivity and sample concordance is variable, highlighting the importance of evaluations in setting of intended use. This rigorous approach to LFIA evaluation identified a test with high specificity (98.6% (95%CI [97.1, 99.4])), moderate sensitivity (84.4% with fingerprick (95%CI [70.5, 93.5])), and moderate concordance, suitable for seroprevalence surveys.
Accurate antibody tests are essential to monitor the SARS-CoV-2 pandemic. Lateral flow immunoassays (LFIAs) can deliver testing at scale. However, reported performance varies, and sensitivity analyses have generally been conducted on serum from hospitalised patients. For use in community testing, evaluation of finger-prick self-tests, in non-hospitalised individuals, is required.
Methods
Sensitivity analysis was conducted on 276 non-hospitalised participants. All had tested positive for SARS-CoV-2 by RT-PCR and were ≥21d from symptom-onset. In phase I we evaluated five LFIAs in clinic (with finger-prick) and laboratory (with blood and sera) in comparison to a) PCR-confirmed infection and b) presence of SARS-CoV-2 antibodies on two “in-house” ELISAs. Specificity analysis was performed on 500 pre-pandemic sera. In phase II, six additional LFIAs were assessed with serum.
Findings
95% (95%CI [92.2, 97.3]) of the infected cohort had detectable antibodies on at least one ELISA. LFIA sensitivity was variable, but significantly inferior to ELISA in 8/11 assessed. Of LFIAs assessed in both clinic and laboratory, finger-prick self-test sensitivity varied from 21%-92% vs PCR-confirmed cases and 22%-96% vs composite ELISA positives. Concordance between finger-prick and serum testing was at best moderate (kappa 0.56) and, at worst, slight (kappa 0.13). All LFIAs had high specificity (97.2% - 99.8%).
Interpretation
LFIA sensitivity and sample concordance is variable, highlighting the importance of evaluations in setting of intended use. This rigorous approach to LFIA evaluation identified a test with high specificity (98.6% (95%CI [97.1, 99.4])), moderate sensitivity (84.4% with fingerprick (95%CI [70.5, 93.5])), and moderate concordance, suitable for seroprevalence surveys.
Date Issued
2020-08-12
Date Acceptance
2020-07-25
Citation
Thorax, 2020, 75 (12), pp.1082-1088
ISSN
0040-6376
Publisher
BMJ Publishing Group
Start Page
1082
End Page
1088
Journal / Book Title
Thorax
Volume
75
Issue
12
Copyright Statement
© Author(s) (or their employer(s)) 2020. No commercial re-use. See rights and permissions. Published by BMJ.
This article is made freely available for use in accordance with BMJ’s website terms and conditions for the duration of the covid-19 pandemic or until otherwise determined by BMJ. You may use, download and print the article for any lawful, non-commercial purpose (including text and data mining) provided that all copyright notices and trade marks are retained.
This article is made freely available for use in accordance with BMJ’s website terms and conditions for the duration of the covid-19 pandemic or until otherwise determined by BMJ. You may use, download and print the article for any lawful, non-commercial purpose (including text and data mining) provided that all copyright notices and trade marks are retained.
License URL
Sponsor
National Institute of Health Research
National Institute for Health Research
Identifier
https://thorax.bmj.com/content/75/12/1082
Grant Number
RP-2016-07-012
Subjects
clinical epidemiology
respiratory infection
viral infection
1103 Clinical Sciences
Respiratory System
Publication Status
Published
Date Publish Online
2020-08-12