Physiological assessment of ventricular myocardial voltage using omnipolar electrograms
Author(s)
Type
Journal Article
Abstract
Background
Characterization of myocardial health by bipolar electrograms are critical for ventricular tachycardia therapy. Dependence of bipolar electrograms on electrode orientation may reduce reliability of voltage assessment along the plane of arrhythmic myocardial substrate. Hence, we sought to evaluate voltage assessment from orientation‐independent omnipolar electrograms.
Methods and Results
We mapped the ventricular epicardium of 5 isolated hearts from each species—healthy rabbits, healthy pigs, and diseased humans—under paced conditions. We derived bipolar electrograms and voltage peak‐to‐peak (Vpps) along 2 bipolar electrode orientations (horizontal and vertical). We derived omnipolar electrograms and Vpps using omnipolar electrogram methodology. Voltage maps were created for both bipoles and omnipole. Electrode orientation affects the bipolar voltage map with an average absolute difference between horizontal and vertical of 0.25±0.18 mV in humans. Vpps provide larger absolute values than horizontal and vertical bipolar Vpps by 1.6 and 1.4 mV, respectively, in humans. Bipolar electrograms with the largest Vpps from either along horizontal or vertical orientation are highly correlated with omnipolar electrograms and with Vpps values (0.97±0.08 and 0.94±0.08, respectively). Vpps values are more consistent than bipoles, in both beat‐by‐beat (CoV, 0.28±0.19 versus 0.08±0.13 in human hearts) and rhythm changes (0.55±0.21 versus 0.40±0.20 in porcine hearts).
Conclusions
Omnipoles provide physiologically relevant and consistent voltages that are along the maximal bipolar direction on the plane of the myocardium.
Characterization of myocardial health by bipolar electrograms are critical for ventricular tachycardia therapy. Dependence of bipolar electrograms on electrode orientation may reduce reliability of voltage assessment along the plane of arrhythmic myocardial substrate. Hence, we sought to evaluate voltage assessment from orientation‐independent omnipolar electrograms.
Methods and Results
We mapped the ventricular epicardium of 5 isolated hearts from each species—healthy rabbits, healthy pigs, and diseased humans—under paced conditions. We derived bipolar electrograms and voltage peak‐to‐peak (Vpps) along 2 bipolar electrode orientations (horizontal and vertical). We derived omnipolar electrograms and Vpps using omnipolar electrogram methodology. Voltage maps were created for both bipoles and omnipole. Electrode orientation affects the bipolar voltage map with an average absolute difference between horizontal and vertical of 0.25±0.18 mV in humans. Vpps provide larger absolute values than horizontal and vertical bipolar Vpps by 1.6 and 1.4 mV, respectively, in humans. Bipolar electrograms with the largest Vpps from either along horizontal or vertical orientation are highly correlated with omnipolar electrograms and with Vpps values (0.97±0.08 and 0.94±0.08, respectively). Vpps values are more consistent than bipoles, in both beat‐by‐beat (CoV, 0.28±0.19 versus 0.08±0.13 in human hearts) and rhythm changes (0.55±0.21 versus 0.40±0.20 in porcine hearts).
Conclusions
Omnipoles provide physiologically relevant and consistent voltages that are along the maximal bipolar direction on the plane of the myocardium.
Date Issued
2017-08-01
Date Acceptance
2017-06-21
Citation
Journal of the American Heart Association, 2017, 6 (8), pp.1-14
ISSN
2047-9980
Publisher
Wiley
Start Page
1
End Page
14
Journal / Book Title
Journal of the American Heart Association
Volume
6
Issue
8
Copyright Statement
© 2017 The Authors. Published on behalf of the American Heart Association, Inc., by Wiley.
This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000427296800044&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Cardiac & Cardiovascular Systems
Cardiovascular System & Cardiology
electrophysiology mapping
omnipole
physiology
ventricular myocardium
ventricular tachycardia
voltage mapping
TACHYCARDIA ABLATION
SCAR
FEASIBILITY
INFARCTION
THRESHOLD
MODEL
Publication Status
Published
Article Number
ARTN e006447
Date Publish Online
2017-08-16