Developing a soft inflatable endorectal actuator for prostate radiotherapy
File(s)
Author(s)
Niknam Maleki, Aryan
Type
Thesis or dissertation
Abstract
Prostate motion during radiotherapy—primarily caused by patient-specific variations in rectal filling—can compromise dose delivery accuracy, increasing the risk of biochemical recurrence and toxicity. Current rectal stabilisation devices suffer from poor tolerability or limited validation. This thesis aimed to design, fabricate, and validate a novel rectal stabilisation device—an endorectal actuator—addressing these limitations.
An initial trial comparing CT scans of patients who could and could not tolerate ProSpare™ found no significant anatomical differences (anorectal angle, rectal diameter, anal canal length), suggesting that reduced tolerability was caused by device geometry and rigidity rather than a mismatch to anorectal anatomy. These results helped establish the seven design criteria: deployability, radial stiffness, rectal retraction capability, self-insertion, image guidance, and low-cost manufacturing.
A hydraulically inflatable soft robotic actuator was developed through successive design iterations to improve device rigidity, control, and durability. Replacing conical pouch motors with elliptical pouch motor reduced internal membrane damage while increasing volume capacity and holding force. A dual-flexor configuration further improved resistance to opposing bending action. The final device achieved a baseline angle of 115.8°, rising to 122.8° under 0.72 N of opposing force.
Radiopaque markers for image guidance proved too thin to be visible on CT, but the water-inflated device itself was visible contrasting with air; a low-cost contrast-enhancing liquid may be viable.
Rectal phantom validation demonstrated the device’s ability to reduce diameter change (from 25 mm to <5 mm) and rectal wall displacement (from 15 mm to <5 mm) under dynamic compressions; and reduce rectal wall displacement (13 mm to 4 mm) and anorectal angle (+12° to +2°) after inducing anterior translation.
The final endorectal actuator met six of seven design criteria, suggesting it may reduce prostate displacement and improve radiotherapy outcomes. Future work should improve usability, soften device edges, and integrate sensing for treatment monitoring.
An initial trial comparing CT scans of patients who could and could not tolerate ProSpare™ found no significant anatomical differences (anorectal angle, rectal diameter, anal canal length), suggesting that reduced tolerability was caused by device geometry and rigidity rather than a mismatch to anorectal anatomy. These results helped establish the seven design criteria: deployability, radial stiffness, rectal retraction capability, self-insertion, image guidance, and low-cost manufacturing.
A hydraulically inflatable soft robotic actuator was developed through successive design iterations to improve device rigidity, control, and durability. Replacing conical pouch motors with elliptical pouch motor reduced internal membrane damage while increasing volume capacity and holding force. A dual-flexor configuration further improved resistance to opposing bending action. The final device achieved a baseline angle of 115.8°, rising to 122.8° under 0.72 N of opposing force.
Radiopaque markers for image guidance proved too thin to be visible on CT, but the water-inflated device itself was visible contrasting with air; a low-cost contrast-enhancing liquid may be viable.
Rectal phantom validation demonstrated the device’s ability to reduce diameter change (from 25 mm to <5 mm) and rectal wall displacement (from 15 mm to <5 mm) under dynamic compressions; and reduce rectal wall displacement (13 mm to 4 mm) and anorectal angle (+12° to +2°) after inducing anterior translation.
The final endorectal actuator met six of seven design criteria, suggesting it may reduce prostate displacement and improve radiotherapy outcomes. Future work should improve usability, soften device edges, and integrate sensing for treatment monitoring.
Version
Open Access
Date Issued
2025-07-02
Date Awarded
2026-08-01
Copyright Statement
Attribution-NonCommercial 4.0 International Licence (CC BY-NC)
License URL
Advisor
Mylonas, George
Murray, Julia
Thompson, Alexander
Jones, Julian
Publisher Department
Department of Surgery & Cancer
Publisher Institution
Imperial College London
Qualification Level
Doctoral
Qualification Name