Obtaining patient torso geometry for the design of scoliosis braces. A study of the accuracy and repeatability of handheld 3D scanners
File(s)POI-D-21-00114_R1 (2)_4-30.pdf (1.81 MB)
Accepted version
Author(s)
Type
Journal Article
Abstract
Objective:
Obtaining patient geometry is crucial in scoliosis brace design for patients with adolescent idiopathic scoliosis. Advances in 3D scanning technologies provide the opportunity to obtain patient geometries quickly with fewer resources during the design process compared with the plaster-cast method. This study assesses the accuracy and repeatability of such technologies for this application.
Methods:
The accuracy and repeatability of three different handheld scanners and phone-photogrammetry was assessed using different mesh generation software. Twenty-four scans of a single subject's torso were analyzed for accuracy and repeatability based on anatomical landmark distances and surface deviation maps.
Results:
Mark II and Structure ST01 scanners showed maximum mean surface deviations of 1.74 ± 3.63 mm and 1.64 ± 3.06 mm, respectively. Deviations were lower for the Peel 1 scanner (maximum of −0.35 ± 2.8 mm) but higher with the use of phone-photogrammetry (maximum of −5.1 ± 4.8 mm). The mean absolute errors of anatomical landmark distance measurements from torso meshes obtained with the Peel 1, Mark II, and ST01 scanners were all within 9.3 mm (3.6%), whereas phone-photogrammetry errors were as high as 18 mm (7%).
Conclusions:
Low-cost Mark II and ST01 scanners are recommended for obtaining torso geometries because of their accuracy and repeatability. Subject’s breathing/movement affects the resultant geometry around the abdominal and anterolateral regions.
Obtaining patient geometry is crucial in scoliosis brace design for patients with adolescent idiopathic scoliosis. Advances in 3D scanning technologies provide the opportunity to obtain patient geometries quickly with fewer resources during the design process compared with the plaster-cast method. This study assesses the accuracy and repeatability of such technologies for this application.
Methods:
The accuracy and repeatability of three different handheld scanners and phone-photogrammetry was assessed using different mesh generation software. Twenty-four scans of a single subject's torso were analyzed for accuracy and repeatability based on anatomical landmark distances and surface deviation maps.
Results:
Mark II and Structure ST01 scanners showed maximum mean surface deviations of 1.74 ± 3.63 mm and 1.64 ± 3.06 mm, respectively. Deviations were lower for the Peel 1 scanner (maximum of −0.35 ± 2.8 mm) but higher with the use of phone-photogrammetry (maximum of −5.1 ± 4.8 mm). The mean absolute errors of anatomical landmark distance measurements from torso meshes obtained with the Peel 1, Mark II, and ST01 scanners were all within 9.3 mm (3.6%), whereas phone-photogrammetry errors were as high as 18 mm (7%).
Conclusions:
Low-cost Mark II and ST01 scanners are recommended for obtaining torso geometries because of their accuracy and repeatability. Subject’s breathing/movement affects the resultant geometry around the abdominal and anterolateral regions.
Date Issued
2022-08-01
Date Acceptance
2022-01-06
Citation
Prosthetics and Orthotics International, 2022, 46 (4), pp.e374-e382
ISSN
0309-3646
Publisher
Lippincott, Williams & Wilkins
Start Page
e374
End Page
e382
Journal / Book Title
Prosthetics and Orthotics International
Volume
46
Issue
4
Copyright Statement
© 2022 Lippincott Williams & Wilkins, Inc. This is a non-final version of an article published in final form in Prosthetics and Orthotics International: March 24, 2022 - Volume - Issue - 10.1097/PXR.0000000000000118
Identifier
https://journals.lww.com/poijournal/Abstract/9900/Obtaining_Patient_Torso_Geometry_for_the_Design_of.15.aspx
Subjects
Rehabilitation
0903 Biomedical Engineering
1106 Human Movement and Sports Sciences
Publication Status
Published
Date Publish Online
2022-03-24