Evaluation of 4D Flow MRI-based non-invasive pressure assessment in aortic coarctations
File(s)SaittaS_JBiomech_Final_withFigures.pdf (1.44 MB)
Accepted version
Author(s)
Type
Journal Article
Abstract
Severity of aortic coarctation (CoA) is currently assessed by estimating trans-coarctation pressure drops through cardiac catheterization or echocardiography. In principle, more detailed information could be obtained non-invasively based on space- and time-resolved magnetic resonance imaging (4D flow) data. Yet the limitations of this imaging technique require testing the accuracy of 4D flow-derived hemodynamic quantities against other methodologies.
With the objective of assessing the feasibility and accuracy of this non-invasive method to support the clinical diagnosis of CoA, we developed an algorithm (4DF-FEPPE) to obtain relative pressure distributions from 4D flow data by solving the Poisson pressure equation. 4DF-FEPPE was tested against results from a patient-specific fluid-structure interaction (FSI) simulation, whose patient-specific boundary conditions were prescribed based on 4D flow data. Since numerical simulations provide noise-free pressure fields on fine spatial and temporal scales, our analysis allowed to assess the uncertainties related to 4D flow noise and limited resolution.
4DF-FEPPE and FSI results were compared on a series of cross-sections along the aorta. Bland-Altman analysis revealed very good agreement between the two methodologies in terms of instantaneous data at peak systole, end-diastole and time-averaged values: biases (means of differences) were +0.4 mmHg, −1.1 mmHg and +0.6 mmHg, respectively. Limits of agreement (2 SD) were ±0.978 mmHg, ±1.06 mmHg and ±1.97 mmHg, respectively. Peak-to-peak and maximum trans-coarctation pressure drops obtained with 4DF-FEPPE differed from FSI results by 0.75 mmHg and −1.34 mmHg respectively. The present study considers important validation aspects of non-invasive pressure difference estimation based on 4D flow MRI, showing the potential of this technology to be more broadly applied to the clinical practice.
With the objective of assessing the feasibility and accuracy of this non-invasive method to support the clinical diagnosis of CoA, we developed an algorithm (4DF-FEPPE) to obtain relative pressure distributions from 4D flow data by solving the Poisson pressure equation. 4DF-FEPPE was tested against results from a patient-specific fluid-structure interaction (FSI) simulation, whose patient-specific boundary conditions were prescribed based on 4D flow data. Since numerical simulations provide noise-free pressure fields on fine spatial and temporal scales, our analysis allowed to assess the uncertainties related to 4D flow noise and limited resolution.
4DF-FEPPE and FSI results were compared on a series of cross-sections along the aorta. Bland-Altman analysis revealed very good agreement between the two methodologies in terms of instantaneous data at peak systole, end-diastole and time-averaged values: biases (means of differences) were +0.4 mmHg, −1.1 mmHg and +0.6 mmHg, respectively. Limits of agreement (2 SD) were ±0.978 mmHg, ±1.06 mmHg and ±1.97 mmHg, respectively. Peak-to-peak and maximum trans-coarctation pressure drops obtained with 4DF-FEPPE differed from FSI results by 0.75 mmHg and −1.34 mmHg respectively. The present study considers important validation aspects of non-invasive pressure difference estimation based on 4D flow MRI, showing the potential of this technology to be more broadly applied to the clinical practice.
Date Issued
2019-09-01
Date Acceptance
2019-07-05
Citation
Journal of Biomechanics, 2019, 94, pp.13-21
ISSN
0021-9290
Publisher
Elsevier
Start Page
13
End Page
21
Journal / Book Title
Journal of Biomechanics
Volume
94
Copyright Statement
© YYYY Elsevier Ltd. All rights reserved. This manuscript is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International Licence http://creativecommons.org/licenses/by-nc-nd/4.0/
Sponsor
Commission of the European Communities
Identifier
https://www.sciencedirect.com/science/article/pii/S0021929019304440?via%3Dihub
Grant Number
642458
Subjects
Science & Technology
Life Sciences & Biomedicine
Technology
Biophysics
Engineering, Biomedical
Engineering
Aortic coarctation
4D flow MRI
Pressure Poisson equation
Non-invasive pressure difference estimation
Fluid dynamics
CONGENITAL HEART-DISEASE
PULSE-WAVE VELOCITY
BLOOD-FLOW
STIFFNESS
QUANTIFICATION
GUIDELINES
MANAGEMENT
VESSELS
4D flow MRI
Aortic coarctation
Fluid dynamics
Non-invasive pressure difference estimation
Pressure Poisson equation
Biomedical Engineering
0903 Biomedical Engineering
1106 Human Movement and Sports Sciences
0913 Mechanical Engineering
Publication Status
Published
Date Publish Online
2019-07-09