The effects of optical sensor-tissue separation in endocavitary photoplethysmography
File(s)2018_0612_Patel_PhysiolMeasurement.pdf (2.46 MB)
Accepted version
Author(s)
Patel, Zaibaa
Thaha, Mohamed A
Kyriacou, Panayiotis A
Type
Journal Article
Abstract
Objective: Intestinal anastomotic failure that occurs mainly due to ischaemia is a serious risk in colorectal cancer patients undergoing surgery. Surgeons continue to rely on subjective methods such as visual inspection to assess intestinal viability during surgery and there are no clinical tools to directly monitor viability postoperatively. A dual-wavelength reflectance optical sensor has been developed for continuous and dynamic monitoring of intestinal viability via the intestinal lumen. Maintaining direct contact between the sensor and the inner intestinal wall can be difficult in an intraluminal design, therefore impacting on signal acquisition and quality. This paper investigates the effect of direct contact versus variable distances between the sensor and the tissue surface of the buccal mucosa as a surrogate. Approach: The in vivo study involved 20 healthy volunteers to measure the effect of optical sensor-tissue distances on the ability to acquire photoplethysmography signals and their quality. Signals were acquired from the buccal mucosa at five optical sensor-tissue distances. Main results: Distances between 0 mm (contact) to 5 mm were the most optimal, producing signals of high quality and signal-to-noise ratio, resulting in reliable estimations of the blood oxygen saturation. Distances exceeding 5 mm compromised the acquired signals, and were of poor quality, thereby unreliably estimating the blood oxygen saturation. Significance: The developed optical sensor proved to be reliable for acquiring photoplethysmography signals for cases where distances between the optical sensor-tissue may arise during the assessment of intraluminal intestinal viability.
Date Issued
2018-07-01
Date Acceptance
2018-06-12
Citation
Physiological Measurement, 2018, 39 (7), pp.1-10
ISSN
0967-3334
Publisher
IOP Publishing
Start Page
1
End Page
10
Journal / Book Title
Physiological Measurement
Volume
39
Issue
7
Copyright Statement
© 2018 Institute of Physics and Engineering in Medicine. , This Accepted Manuscript is available for reuseunder a CC BY-NC-ND 3.0 licence (https://creativecommons.org/licenses/by-nc-nd/3.0/)
Identifier
http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000437423900001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Technology
Biophysics
Engineering, Biomedical
Physiology
Engineering
bowel viability
intestine
optical sensor
photoplethysmography
physiological monitoring
pulse oximetry
PULSE OXIMETRY
PERFUSION
Publication Status
Published
Article Number
ARTN 075001
Date Publish Online
2018-07-03