Particle release from dental implants immediately after placement - An ex vivo comparison of different implant systems
File(s)barrak wear particles.pdf (10.16 MB)
Published version
Author(s)
Type
Journal Article
Abstract
Objectives
Metallic element release during implant placement can lead to mucositis and peri-implantitis. Here, using ex vivo porcine mandibles, the release of metallic elements into the surrounding bone with different material and geometrical designs was quantified.
Methods
Implants from BioHorizons® and Straumann® (Bone level, tapered/cylindrical, 3/4 mm body diameter, Ti-CP4/Ti-6Al-4V/Ti-15Zr) systems were used. Micro computed tomography and inductively coupled plasma optical emission spectroscopy was used to visualise and quantify metallic elements in bone, following acid digestion. Implant surfaces were examined with scanning electron microscopy and internalization of implant particles by human gingival fibroblasts (HGFs) and RAW 264.7 macrophages were demonstrated in vitro.
Results
Implants with wider body diameters resulted in higher metallic element release. Ti-6Al-4V implants released significantly more metallic elements in comparison to both Ti-CP4 and Ti-15Zr devices with similar design and dimensions. Tapered Ti-CP4 implants released less compared to those with cylindrical design. Al three types of particles were internalized by HGFs and RAW 264.7.
Significance
Ti-CP4 and Ti-15Zr appear to be more suitable materials, however, further studies are required to elucidate the biological effects of the fine particles and/or metallic species from dental implants. Authors would like to raise the awareness in the dental profession community that careful evaluation of the materials used in dental implants and the potential risks of the individual constituents of any alloy are needed. The potential cytotoxicity of Ti-6Al-4V implant particles should be highlighted. Further investigations on the biological effect of the fine particles or metallic species released from dental implants are also needed.
Metallic element release during implant placement can lead to mucositis and peri-implantitis. Here, using ex vivo porcine mandibles, the release of metallic elements into the surrounding bone with different material and geometrical designs was quantified.
Methods
Implants from BioHorizons® and Straumann® (Bone level, tapered/cylindrical, 3/4 mm body diameter, Ti-CP4/Ti-6Al-4V/Ti-15Zr) systems were used. Micro computed tomography and inductively coupled plasma optical emission spectroscopy was used to visualise and quantify metallic elements in bone, following acid digestion. Implant surfaces were examined with scanning electron microscopy and internalization of implant particles by human gingival fibroblasts (HGFs) and RAW 264.7 macrophages were demonstrated in vitro.
Results
Implants with wider body diameters resulted in higher metallic element release. Ti-6Al-4V implants released significantly more metallic elements in comparison to both Ti-CP4 and Ti-15Zr devices with similar design and dimensions. Tapered Ti-CP4 implants released less compared to those with cylindrical design. Al three types of particles were internalized by HGFs and RAW 264.7.
Significance
Ti-CP4 and Ti-15Zr appear to be more suitable materials, however, further studies are required to elucidate the biological effects of the fine particles and/or metallic species from dental implants. Authors would like to raise the awareness in the dental profession community that careful evaluation of the materials used in dental implants and the potential risks of the individual constituents of any alloy are needed. The potential cytotoxicity of Ti-6Al-4V implant particles should be highlighted. Further investigations on the biological effect of the fine particles or metallic species released from dental implants are also needed.
Date Issued
2022-06-01
Date Acceptance
2022-04-01
Citation
Dental Materials, 2022, 38 (6), pp.1004-1014
ISSN
0109-5641
Publisher
Elsevier
Start Page
1004
End Page
1014
Journal / Book Title
Dental Materials
Volume
38
Issue
6
Copyright Statement
© 2022 The Authors. Published by Elsevier Inc. on behalf of The Academy of Dental Materials.
CC_BY_4.0
CC_BY_4.0
License URL
Identifier
https://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=WOS:000829945500002&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=1ba7043ffcc86c417c072aa74d649202
Subjects
Science & Technology
Life Sciences & Biomedicine
Technology
Dentistry, Oral Surgery & Medicine
Materials Science, Biomaterials
Materials Science
Dental implant
Particle release
Wear
Commercial pure titanium
Titanium alloy
Ti-6Al-4V
Toxicity
Roxolid
Phagocytosis
TITANIUM PARTICLES
INDUCED OSTEOLYSIS
ORAL-MUCOSA
NANOPARTICLES
CORROSION
INSERTION
ALLOYS
BONE
TI
NANOMATERIALS
Publication Status
Published
Date Publish Online
2022-04-21