Real-time tracking of a diffuse reflectance spectroscopy probe used to aid histological validation of margin assessment in upper gastrointestinal cancer resection surgery
File(s) 025001_1.pdf (2.5 MB)
Published version
Author(s)
Type
Journal Article
Abstract
Significance: Diffuse reflectance spectroscopy (DRS) allows discrimination of tissue type. Its application is limited by the inability to mark the scanned tissue and the lack of real-time measurements.
Aim: This study aimed to develop a real-time tracking system to enable localization of a DRS probe to aid the classification of tumor and non-tumor tissue.
Approach: A green-colored marker attached to the DRS probe was detected using hue-saturation-value (HSV) segmentation. A live, augmented view of tracked optical biopsy sites was recorded in real time. Supervised classifiers were evaluated in terms of sensitivity, specificity, and overall accuracy. A developed software was used for data collection, processing, and statistical analysis.
Results: The measured root mean square error (RMSE) of DRS probe tip tracking was 1.18 ± 0.58 mm and 1.05 ± 0.28 mm for the x and y dimensions, respectively. The diagnostic accuracy of the system to classify tumor and non-tumor tissue in real time was 94% for stomach and 96% for the esophagus.
Conclusions: We have successfully developed a real-time tracking and classification system for a DRS probe. When used on stomach and esophageal tissue for tumor detection, the accuracy derived demonstrates the strength and clinical value of the technique to aid margin assessment in cancer resection surgery.
Aim: This study aimed to develop a real-time tracking system to enable localization of a DRS probe to aid the classification of tumor and non-tumor tissue.
Approach: A green-colored marker attached to the DRS probe was detected using hue-saturation-value (HSV) segmentation. A live, augmented view of tracked optical biopsy sites was recorded in real time. Supervised classifiers were evaluated in terms of sensitivity, specificity, and overall accuracy. A developed software was used for data collection, processing, and statistical analysis.
Results: The measured root mean square error (RMSE) of DRS probe tip tracking was 1.18 ± 0.58 mm and 1.05 ± 0.28 mm for the x and y dimensions, respectively. The diagnostic accuracy of the system to classify tumor and non-tumor tissue in real time was 94% for stomach and 96% for the esophagus.
Conclusions: We have successfully developed a real-time tracking and classification system for a DRS probe. When used on stomach and esophageal tissue for tumor detection, the accuracy derived demonstrates the strength and clinical value of the technique to aid margin assessment in cancer resection surgery.
Date Issued
2022-02-01
Date Acceptance
2022-01-10
Citation
Journal of Biomedical Optics, 2022, 27 (2)
ISSN
1083-3668
Publisher
Society of Photo-optical Instrumentation Engineers
Journal / Book Title
Journal of Biomedical Optics
Volume
27
Issue
2
Copyright Statement
© The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
License URL
Sponsor
Cancer Research UK
Imperial College Healthcare NHS Trust- BRC Funding
Grant Number
25147
RDB04
Subjects
cancer
diffuse reflectance spectroscopy
machine learning
margin delineation
probe tracking
tissue classification
Optics
0205 Optical Physics
0903 Biomedical Engineering
1113 Opthalmology and Optometry
Publication Status
Published
Article Number
ARTN 025001