Background
Valved holding chambers (VHCs) and spacers are widely used in clinical practice, but the lack of standardised in vitro methodology makes it difficult to compare device performance and guide evidence-based clinical decisions. We aimed to synthesise the clinically relevant drug delivery features of these devices and map the breadth and quality of evidence supporting their use across age groups.
Methods
We conducted a scoping review with systematic searches on four databases from inception to June 2024, updated February 2026. Eligible studies investigated drug delivery performance of VHCs, spacers, or similar auxiliary inhalation devices using laboratory-based techniques. Two independent reviewers screened and extracted data. Data were narratively synthesised by device features, methodological variability, and reported financial conflicts of interest (COI). Descriptive charts mapped temporal trends in study designs and device types. (OSF registration: PGW3E)
Results
95 studies were identified since 1990: 66 used steady inhalation flow and 38 used simulated or in vivo patient breathing. 82 distinct VHC or spacer models were tested; 46% were assessed in more than one study. Steady inhalation flow testing dominated, and breathing simulation studies frequently relied on filter-based endpoints rather than full particle-size characterisation. Key device descriptors likely to influence performance, including material, electrostatics, and valve configuration, were often unreported, limiting interpretability and reproducibility. 59/95 (62%) of studies either declared a financial COI (37/95, 39%) or provided no statement (22/95, 23%).
Conclusions
Evidence on VHCs and spacers is wide in device coverage but limited in replication per device. Methodological heterogeneity and COI limit comparability. Standardised methodology is needed for evidence-based device selection.
Valved holding chambers (VHCs) and spacers are widely used in clinical practice, but the lack of standardised in vitro methodology makes it difficult to compare device performance and guide evidence-based clinical decisions. We aimed to synthesise the clinically relevant drug delivery features of these devices and map the breadth and quality of evidence supporting their use across age groups.
Methods
We conducted a scoping review with systematic searches on four databases from inception to June 2024, updated February 2026. Eligible studies investigated drug delivery performance of VHCs, spacers, or similar auxiliary inhalation devices using laboratory-based techniques. Two independent reviewers screened and extracted data. Data were narratively synthesised by device features, methodological variability, and reported financial conflicts of interest (COI). Descriptive charts mapped temporal trends in study designs and device types. (OSF registration: PGW3E)
Results
95 studies were identified since 1990: 66 used steady inhalation flow and 38 used simulated or in vivo patient breathing. 82 distinct VHC or spacer models were tested; 46% were assessed in more than one study. Steady inhalation flow testing dominated, and breathing simulation studies frequently relied on filter-based endpoints rather than full particle-size characterisation. Key device descriptors likely to influence performance, including material, electrostatics, and valve configuration, were often unreported, limiting interpretability and reproducibility. 59/95 (62%) of studies either declared a financial COI (37/95, 39%) or provided no statement (22/95, 23%).
Conclusions
Evidence on VHCs and spacers is wide in device coverage but limited in replication per device. Methodological heterogeneity and COI limit comparability. Standardised methodology is needed for evidence-based device selection.
Date Acceptance
2026-10-12
Citation
European Respiratory Review
ISSN
0905-9180
Publisher
European Respiratory Society
Journal / Book Title
European Respiratory Review
Copyright Statement
Copyright This paper is embargoed until publication. Once published the author’s accepted manuscript will be made available under a CC-BY License in accordance with Imperial’s Research Publications Open Access policy (www.imperial.ac.uk/oa-policy).
License URL
Publication Status
Accepted