Comparing HemoCue® and Quantitative Buffy Coat® and Coulter Counter-measured haemoglobin concentrations in African children with acute uncomplicated malaria: a Bland–Altman analysis
File(s) s12936-025-05318-5.pdf (1.34 MB)
Published version
Author(s)
Type
Journal Article
Abstract
Background
Anaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland–Altman analysis.
Methods
As part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0, 3, 7, and 28 using Coulter Counter (Uganda, n = 1880 paired values), QBC® (DR Congo, n = 1984 paired values) and HemoCue® Hb-301™. The predefined clinically acceptable limits were set at ± 0.5 g/dL.
Results
The Bland–Altman analysis showed that the HemoCue® minus Coulter Counter mean Hb difference was − 0.15 g/dL with lower and upper limits of agreement of − 3.68 g/dL and 3.39 g/dL, respectively. Corresponding HemoCue® minus QBC® values were − 0.23 g/dL, − 1.66 g/dL and 1.22 g/dL. Linear regression of Hb concentration differences vs. mean Hb concentrations showed negative correlations: r = − 0.43 and r = − 0.34 for HemoCue® vs. Coulter Counter and HemoCue® vs. QBC®, respectively.
Conclusions
Compared to Coulter and QBC®, mean HemoCue® measured Hb concentrations were lower and, compared to the Coulter or QBC® methods, had an overall tendency to measure lower Hb concentrations with increasing Hb concentrations. Upper and lower limits of agreement were wider than the predefined clinically acceptable limits of ± 0.5 g/dL. HemoCue® should be used with caution in settings where decisions about blood transfusions are made.
Anaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland–Altman analysis.
Methods
As part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0, 3, 7, and 28 using Coulter Counter (Uganda, n = 1880 paired values), QBC® (DR Congo, n = 1984 paired values) and HemoCue® Hb-301™. The predefined clinically acceptable limits were set at ± 0.5 g/dL.
Results
The Bland–Altman analysis showed that the HemoCue® minus Coulter Counter mean Hb difference was − 0.15 g/dL with lower and upper limits of agreement of − 3.68 g/dL and 3.39 g/dL, respectively. Corresponding HemoCue® minus QBC® values were − 0.23 g/dL, − 1.66 g/dL and 1.22 g/dL. Linear regression of Hb concentration differences vs. mean Hb concentrations showed negative correlations: r = − 0.43 and r = − 0.34 for HemoCue® vs. Coulter Counter and HemoCue® vs. QBC®, respectively.
Conclusions
Compared to Coulter and QBC®, mean HemoCue® measured Hb concentrations were lower and, compared to the Coulter or QBC® methods, had an overall tendency to measure lower Hb concentrations with increasing Hb concentrations. Upper and lower limits of agreement were wider than the predefined clinically acceptable limits of ± 0.5 g/dL. HemoCue® should be used with caution in settings where decisions about blood transfusions are made.
Date Issued
2025-03-11
Date Acceptance
2025-03-03
Citation
Malaria Journal, 2025, 24
ISSN
1475-2875
Publisher
Springer Science and Business Media LLC
Journal / Book Title
Malaria Journal
Volume
24
Copyright Statement
© The Author(s) 2025. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
License URL
Identifier
10.1186/s12936-025-05318-5
Subjects
Bland-Altman analysis
Haemoglobin
Anaemia
Malaria
HemoCue
Coulter Counter
QBC ® †